Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "MiPNet14.06 Instrumental O2 background"

From Bioblast
 
(124 intermediate revisions by 24 users not shown)
Line 1: Line 1:
{{Template:OROBOROS support page name}}
{{Publication
{{Publication
|title=Fasching M, Gnaiger E (2009) Instrumental background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6.
|title=[[Image:O2k-Manual.jpg|right|70px|link=O2k-Manual|O2k-Manual]] O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux.
|info=[[Media:MiPNet14.06 InstrumentalBackground.pdf|'''Open Access-MiPNet14.06.pdf''']]; [http://www.bioblast.at/index.php/File:MiPNet14.06_InstrumentalBackground.pdf Versions]  [[Image:O2k-Protocols.jpg|right|150px|link=http://www.oroboros.at/?o2k-protocols|O2k-Protocols contents]]
|info=[[File:PDF.jpg|100px|link=http://wiki.oroboros.at/images/6/65/MiPNet14.06_InstrumentalO2Background.pdf |Bioblast pdf]] » [http://www.bioblast.at/index.php/File:MiPNet14.06_InstrumentalO2Background.pdf Versions]
|authors=OROBOROS
|authors=Oroboros
|year=2009
|year=2023-10-19
|journal=Mitochondr Physiol Network
|journal=Mitochondr Physiol Network
|abstract=Correction for the instrumental background is a new and essential standard in high resolution respirometry, automatically performed by [http://www.oroboros.at/index.php?id=datlab DatLab]. Background measurements provide a test of instrument function. In the OROBOROS Oxygraph-2k, background corrections are usually within a few % of experimental flux over the entire experimental oxygen range. At minimum activities, however, even the small background effects become highly significant and require routine correction.
|abstract=Timón-Gómez A, Grings M, Baglivo E, Schmitt S, Gnaiger E (2023) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(09):1-16.
{{MiPNet pdf page linking to MitoPedia}}
:» Product: [[O2k-FluoRespirometer]], [[Oroboros O2k-Catalogue |O2k-Catalogue]]
|mipnetlab=AT_Innsbruck_Oroboros
}}


[[Image:MiPNet14.06.jpg|centre|frame|Instrumental background experiment, measuring oxygen flux without biological sample at four oxygen levels (left), and linear relation between instrumental background oxygen flux and oxygen concentration (right). Modified after: Gnaiger E. (2001).]]
== Keywords ==
{{Template:Keywords: Chamber volume}}


Autoxidation of ascorbate and TMPD causes a [http://bioblast.at/images/0/0b/MiPNet06.06_ChemicalBackground.pdf chemical background oxygen flux]. DatLab provides on-line correction for instrumental and chemical background.


[http://www.oroboros.at/index.php?id=backgroundcorrection#c3005 Excel Templates and DatLab-Demo Files]




'''O2k-Protocol for Instrumental Background Correction and Accuracy of Oxygen Flux '''
'''Acknowledgements'''
[[File:Template NextGen-O2k.jpg|left|400px|link=NextGen-O2k]]
<br/>
<br/>
<br/>
<br/>


For calibration of the polarographic oxygen sensor (POS) and for measurement of instrumental background oxygen consumption, only incubation medium but no biological sample is added into the Oxygraph-2k chamber, at experimental conditions. In a closed chamber under these conditions, ideally oxygen concentration remains constant. In practice, however, instrumental background effects are caused by back-diffusion into the oxygraph chamber at low oxygen pressure, oxygen diffusion out of the oxygraph chamber at elevated oxygen levels, and oxygen consumption by the polarographic oxygen sensor (POS). Instrumental background interferes with accurate measurement of respiratory oxygen flux, if background effects remain undefined. Determination of instrumental background constitutes an important standard operating procedure (SOP) in high-resolution respirometry (HRR) Instrumental background oxygen flux is (i) minimized in the OROBOROS Oxygraph-2k by instrumental design and selection of appropriate materials. In addition, (ii) instrumenal background is routinely tested, and (iii) background correction of oxygen flux is applied automatically by DatLab.
:>> O2k-Protocols:[http://www.oroboros.at/?o2k-protocols Overall contents]
:>> Product: [http://www.oroboros.at/?oxygraph OROBOROS Oxygraph-2k], [[O2k-Catalogue_OROBOROS| O2k-Catalogue]]
|keywords=Instrumental background
|mipnetlab=AT_Innsbruck_OROBOROS
|discipline=Mitochondrial Physiology
|articletype=Protocol; Manual, MiPNet-online Publication
}}
{{Labeling
{{Labeling
|area=Respiration, Instruments;methods
|area=Respiration, Instruments;methods
|instruments=Oxygraph-2k, Protocol
|instruments=Oxygraph-2k, O2k-Manual
|additional=O2k-SOP, DatLab
|additional=DatLab, O2k-SOP, DL7, System, MitoFit2022QC
|discipline=Mitochondrial Physiology
|articletype=Protocol; Manual, MiPNet-online Publication
}}
}}
== Instrumental background test for permeabilized muscle fibres ==
=== Extended instrumental background test ===
# While biopsy sampling and fibre preparation proceed: Air calibration in [[MiR06Cr]], then close the chamber to evaluate instrumental background at air saturation (c. 10 min): This is a quality control of the medium, important under field conditions, where medium preservation (sterility) may be less controlled than in the lab.
# Elevate oxygen concentration to 450 µM with oxygen gas ([[Syringe\60 ml\Gas-Injection]]), close and after two to three min perform a [[stirrer test]] (the new DatLab has an automatic stirrer test function, with defined duriation of stopping the stirrer and automatic restart of stirring, 40 s stirrer stop may be optimum). This is important, since the [[OroboPOS]] may have a different [[response time]] at elevated oxygen concentration. If the response time increases dramatically, then the sensor may even show a non-linear response to oxygen concentration at high oxygen levels.
# Instrumental background: After 20 min, open the chamber and allow O<sub>2</sub> to drop to c. 350 µM, close for 20 min, open and drop O<sub>2</sub> to c. 250 µM (this should be the lowest experimental O<sub>2</sub> concentration).
# Increase O<sub>2</sub> with H<sub>2</sub>O<sub>2</sub> injection (c. 2 µl) to 400 µM, measure for 15-20 min instrumental background, simulating a re-oxygenation during the experiment.
# Increase O<sub>2</sub> with H<sub>2</sub>O<sub>2</sub> injection (c. 1 µl) to 450 µM, until the fibres are added, for equilibrating the instrument at high O<sub>2</sub>.
# Addition of permeabilized fibres into the O2k-Chamber: >> [[Permeabilized muscle fibres]]
== Further information ==
'''References'''
[[Gnaiger_1995_J_Bioenerg_Biomembr| Gnaiger E, Steinlechner-Maran R, Méndez G, Eberl T, Margreiter R (1995) Control of mitochondrial and cellular respiration by oxygen. J Bioenerg Biomembr 27: 583-596.]]
[[Gnaiger_2001_Respir_Physiol| Gnaiger E (2001) Bioenergetics at low oxygen: dependence of respiration and phosphorylation on oxygen and adenosine diphosphate supply. Respir Physiol 128: 277-297.]]
[[Gnaiger_2008_POS| Gnaiger E (2008) Polarographic oxygen sensors, the oxygraph and high-resolution respirometry to assess mitochondrial function. In: Mitochondrial Dysfunction in Drug-Induced Toxicity (Dykens JA, Will Y, eds) John Wiley: 327-352.]]
:>> [http://www.oroboros.at/?BackgroundCorrection Intrumental '''Excel Templates and DatLab-Demo Files''']
:>> MiPNet10.07 is integrated in MiPNet14.06.
* Version 2: 2011-12-04
* Version 1: 2009
[[Category:OroboPedia]]

Latest revision as of 17:19, 19 October 2023


                  


O2k-Open Support

MiPNet14.06 Instrumental O2 background


Publications in the MiPMap
O2k-Manual
O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux.

» Bioblast pdf » Versions

Oroboros (2023-10-19) Mitochondr Physiol Network

Abstract: Timón-Gómez A, Grings M, Baglivo E, Schmitt S, Gnaiger E (2023) O2k Quality Control 2: Instrumental oxygen background correction and accuracy of oxygen flux. Mitochondr Physiol Network 14.6(09):1-16.

O2k-technical support and open innovation
Open the pdf document above.
» Current O2k-series: NextGen-O2k Series XB and O2k Series J
» Current software versions DatLab 8.0: MitoPedia: DatLab
» Product: O2k-FluoRespirometer, O2k-Catalogue


O2k-Network Lab: AT_Innsbruck_Oroboros


Keywords


Questions.jpg


Click to expand or collaps




Acknowledgements

Template NextGen-O2k.jpg






Labels: MiParea: Respiration, Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual 

DatLab, O2k-SOP, DL7, System, MitoFit2022QC