Template:Fluorescence dyes and settings: Difference between revisions
From Bioblast
No edit summary |
mNo edit summary |
||
Line 3: | Line 3: | ||
! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage)Β !! Gain !! Comment | ! Application !!Sensor!! Filter set !! O2k output !! Light intensity (Amp polarization voltage)Β !! Gain !! Comment | ||
|- Β | |- Β | ||
| [[Amplex UltraRed|Amplex<sup>TM</sup> UltraRed]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type II]] || 100 - 500 || 1000 || at c(AmR) = 10 Β΅M | | [[Amplex UltraRed|Amplex<sup>TM</sup> UltraRed]]||[[Fluorescence-Sensor Green]] [[Smart Fluo-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type II]] || 100 - 500 || 1000 || at c(AmR) = 10 Β΅M | ||
|- Β | |- Β | ||
| [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 100 - 500|| 1000 || at c(TMRM) = 2 Β΅M | | [[TMRM]]||[[Fluorescence-Sensor Green]] || [[ Filter Set AmR| AmR]] || [[O2k_signal_and_output|Type III]] || 100 - 500|| 1000 || at c(TMRM) = 2 Β΅M |
Revision as of 10:22, 19 November 2021
Application | Sensor | Filter set | O2k output | Light intensity (Amp polarization voltage) | Gain | Comment |
---|---|---|---|---|---|---|
AmplexTM UltraRed | Fluorescence-Sensor Green Smart Fluo-Sensor Green | AmR | Type II | 100 - 500 | 1000 | at c(AmR) = 10 Β΅M |
TMRM | Fluorescence-Sensor Green | AmR | Type III | 100 - 500 | 1000 | at c(TMRM) = 2 Β΅M |
Safranin | Fluorescence-Sensor Blue | Saf | Type III | 100 - 250 | 1000 | at c(safranin) = 2 Β΅M |
Rhodamine 123 | Fluorescence-Sensor Blue | Saf | Type III | 100 - 250 | 1000 | at c(Rh123) = 1 Β΅M |
Magnesium Greenβ’ | Fluorescence-Sensor Blue | MgG / CaG | Type I and II | 100 - 500 | 1000 | at c(Mg Green) = 2 Β΅M |
Calcium Greenβ’ | Fluorescence-Sensor Blue | MgG / CaG | Type I | 300 - 500 | 1000 | at c(Ca Green) = 2 Β΅M |
In the O2k-LED2-Module (Series D-G) polarization voltage is used, while in the O2k-Fluo Smart Module (Series H-I) fluorescence intensity. To use the same setting in the O2k-Fluo LED2-Module as in the O2k-Fluo Smart-Module use the following Excel sheet:
Download the EXCEL calculation-sheet to adjust fluorescence intensity |