::::[[User:Antunes Diana|Antunes Diana]] ([[User talk:Antunes Diana|talk]]) 08:58, 5 May 2020 (CEST) Format for new Agenda/Alerts entries:
<pre style="display: inline-block;">
{|
| logos
| title in bold <br>
- link
|}
</pre>
''e.g.,''
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]
|'''Did you know that the proper use of TIP2k-Filter Papers may avoid an unstable O2 slope neg. signal during Instrumental background correction?''' <br>
- »[[Oxygen_flux_-_instrumental_background#Unstable_O2_slope_neg._signal_in_O2_instrumental_BG|Instrumental background]]« and »[[TIP2k Filter Papers]]«
|}
:::: ''See the scheme'':
<br>
[[Image:TEMPLATE ALERT.png|none|900px|Agenda/Alerts body text format]]
== MitoPedia alert ==
::: Example:
:{|
| width="110" align="center" | [[File:MitoPedia.jpg|left|80px|MitoPedia|link=MitoPedia]]
|''' Measuring fatty acid oxidation (FAO)-linked respiration requires a low concentration of malate '''<br>
In the absence of malate, accumulation of Acetyl-CoA inhibits FAO. <br>
- »'''[[Fatty acid oxidation pathway control state]]'''«, »'''[https://www.bioblast.at/index.php/Gnaiger_2020_BEC_MitoPathways Mitochondrial Pathways: Chapter 5]'''« <br>
- communicated by »[[Cecatto Cristiane]]« and »[[Cardoso Luiza]]«
|}


== Bioblast alert 2020 ==


=== Next ===
=== Next ===


:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|40px|Info: improve]] '''Do you know how to avoid underestimating/overestimating your respiratory fluxes? Measuring steady states with high-resolution respirometry is the key to obtaining correct values.'''
::::::: Learn more about steady states measured with high-resolution respirometry: performance and mark setting for data analysis - »[[Steady state]]«


::::[[Image:O2k-Publications.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]] Parkinson’s Disease Awareness Month:''' Dominant mutation in Parkinson’s Disease causes loss of protein function in mitochondria'''
=== Under construction ===
::::::: - from the O2k-Network »[[FR Angers Gueguen N]]« and »[[BE Leuven Spinazzi M]]«
::::::* Cornelissen T, Spinazzi M, Martin S, Imberechts D, Vangheluwe P, Bird M, De Strooper B, Vandenberghe W (2020) CHCHD2 harboring the Parkinson's disease-linked T61I mutation precipitates inside mitochondria and induces precipitation of wild-type CHCHD2. Hum Mol Genet [Epub ahead of print]. [[Cornelissen 2020 Hum Mol Genet|»Bioblast link«]]


on hold until the page is updated
:{|
| width="100" align="center" |[[File:FAT4BRAIN logo.png|right|80px|link=https://wiki.oroboros.at/index.php/FAT4BRAIN|FAT4BRAIN]]
|'''A FAO-mito-CNS platform to help fatty acid-related respiration measurements was developed in the FAT4BRAIN project''' <br>
- » [[FAT4BRAIN_FAO-mito-CNS_platform|FAT4BRAIN FAO-mito-CNS platform]] «
|}


:::: [[File:O2k-brief.png|45px|link=http://wiki.oroboros.at/index.php/O2k-brief|O2k-brief repository]] '''Novel diurnal action of of mitochondrial stress-induced GDF15 in systemic metabolism regulation'''
== O2k-Brief alert ==
:::* Example:
:{|
| width="100" align="center" | [[File:O2k-brief.png|45px|link=http://wiki.oroboros.at/index.php/O2k-brief|O2k-brief repository]]
|'''Antitumoral properties of triterpenic acid functionalized gold nanoparticles arise from decreasing mitochondrial respiration''' <br>
* Mioc M, Mioc A, Racoviceanu R, Ghiulai R, Prodea A, Milan A, Barbu Tudoran L, Oprean C, Ivan V, Șoica C (2023) The antimelanoma biological assessment of triterpenic acid functionalized gold nanoparticles.  [[Mioc 2023 Molecules|»Bioblast link«]]
|}
=== Next ===


::::::: - from the O2k-Network »[[DE Nuthetal Klaus S]]«
== O2k-Publications alert ==
::::::* Ost M, Igual Gil C, Coleman V, Keipert S, Efstathiou S, Vidic V, Weyers M, Klaus S (2020) Muscle-derived GDF15 drives diurnal anorexia and systemic metabolic remodeling during mitochondrial stress. EMBO Rep [Epub ahead of print].  - [[Ost 2020 EMBO Rep |»Bioblast link«]]
<br />


:::: [[Image:O2k-Network.png|left|70px|link=http://wiki.oroboros.at/index.php/O2k-Network|O2k-Network]] An amazing number of new [[O2k-Publications: Topics |O2k-Publications]] from China » '''[[O2k-Network Labs China]]''' «
:::* Example:  
<br />
:{|
| width="100" align="center" |[[Image:O2k-Publications.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]
|'''"Mitochondria–nucleus communication and its integration within the maintenance of cellular homeostasis constitute key determinants of cellular fate and organism health in the context of cellular stress, homeostasis perturbation, and disease"''' <br>
* Temelie M, Talpur R, Dominguez-Prieto M, Dantas Silva A, Cenusa C, Craciun L, Savu DI, Moisoi N (2023) Impaired integrated stress response and mitochondrial integrity modulate genotoxic stress impact and lower the threshold for immune signalling. [[Temelie 2023 Int J Mol Sci|»Bioblast link«]]
:from the O2k-Network »[[UK Leicester Moisoi N]]«
|}


:::: [[Image:Gentle-Science_Ganesha.jpg|left|60px|link=http://www.bioblast.at/index.php/Gentle_Science|Gentle Science]] '''Let's talk COVID-19 - a new culture of communication in science.''' - [https://www.sciencemag.org/news/2020/02/completely-new-culture-doing-research-coronavirus-outbreak-changes-how-scientists Kay Kupferschmidt in Science]
===='''Author names of O2k-Publications/ Brief for the Agenda vs Bioblast''' ====
<br />


__NOTOC__ [[File:Expand.png|right|25px|Click to expand or collaps]]
<div class="toccolours mw-collapsible mw-collapsed">
'''Guidelines to update a reference'''
<div class="mw-collapsible-content">
::When creating an O2k-Publication/Brief alert, it is the responsibility of the author to check if the reference is accordingly to the correct guidelines:
[[File:Reference example.png|900px]]
::Correct format for a person called “First Name Surname”:
::In the gray box should be “Surname FN” - '''From where references to be used as O2k-Publication/Brief alerts are copied from.'''
[[File:Gray box reference.png|900px]]
::In the links should be “ Surname First N”
[[File:Authors names links reference.png|900px]]
::'''How to edit:'''
:::A) Check if all authors with profiles in Bioblast (blue links) have their profiles with full name. If they have, proceed to next step (B). If not, you need to move it following the steps “How to move a profile” below.
:::B) Click in “Edit with form”.
:::C) The gray box is edited in “Reference” and the links are changed in “Author(s)”. Change it accordingly and click in “save page” at the end of the page.
[[File:Editing reference.png|900px]]


:::: [[Image:MitoFit.jpg|left|50px|link=http://www.mitofit.org/index.php/MitoFit |MitoFit]] The MitoFit Preprint Archives was created as an Open Access preprint server for mitochondrial physiology and bioenergetics - »'''[[MitoFit Preprint Archives]]'''«
'''How to move a profile (instructions sent by Marija):'''
<br />
::::[[Image:O2k-Publications.jpg|60px|left|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]] ''' Zika virus infection leads to mitochondrial dysfunction'''
::::::: - from the O2k-Network »[[BR Rio de Janeiro Galina A]]«
::::::* Ledur PF, Karmirian K, Pedrosa CDSG, Souza LRQ, Assis-de-Lemos G, Martins TM, Ferreira JCCG, de Azevedo Reis GF, Silva ES, Silva D, Salerno JA, Ornelas IM, Devalle S, Madeiro da Costa RF, Goto-Silva L, Higa LM, Melo A, Tanuri A, Chimelli L, Murata MM, Garcez PP, Filippi-Chiela EC, Galina A, Borges HL, Rehen SK (2020) Zika virus infection leads to mitochondrial failure, oxidative stress and DNA damage in human iPSC-derived astrocytes. Sci Rep 10:1218. - [[Ledur 2020 Sci Rep |»Bioblast link«]]


:::: [[File:O2k-brief.png|45px|link=http://wiki.oroboros.at/index.php/O2k-brief|O2k-brief repository]]  '''Could exercise protect against the effects of aging on mitochondrial function?'''
So let’s take for example my page:  
[[Beno M]]
[https://wiki.oroboros.at/index.php?title=Beno_Marija&redirect=no]


::::::: - from the O2k-Network »[[KR Busan Han J]]« »[[KR Incheon Kwak HB]]«
::As you can see there is a redirect from Beno Marija to Beno M. What has to be done is to switch the content of those pages. Here are the steps:
::::::* No MH, Heo JW, Yoo SZ, Kim CJ, Park DH, Kang JH, Seo DY, Han J, Kwak HB (2020) Effects of aging and exercise training on mitochondrial function and apoptosis in the rat heart. Pflugers Arch 472:179-93. - [[No 2020 Pflugers Arch |»Bioblast link«]]
:::1. Go to e.g. “Beno M” on edit
:::2. Click on “Move”
:::3. Change the new title to the full name of the person. In my case it will be Beno M to Beno Marija
[[File:BenoMreference.jpg|500px]]
:::4. Now it should look like that:
[[File:BenoMarijaReference.png|500px]]
:::5. Click on “Move” and leave all the ticked boxes
:::6.Done
::After that you can continue in step B.
::If you have any doubt you can contact Cristiane, Marija or Mario.
</div>
</div>


=== On hold ===
=== Next ===
April 25th


::::[[File:MitoPedia.jpg|60px|MitoPedia]] Did you ever think that units for body mass could be an issue? » '''[[Body mass]]''' «
:{|
| width="100" align="center" |[[Image:O2k-Publications.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]
|'''"Our results further enhance the understanding and relevance of bioenergetic function in liver ischemia and transplantation and provide the basis for further consideration of these parameters as biomarkers in LT [liver transplantation]."''' <br>
* Meszaros AT, Weissenbacher A, Schartner M, Egelseer-Bruendl T, Hermann M, Unterweger J, Mittelberger C, Reyer BA, Hofmann J, Zelger BG, Hautz T, Resch T, Margreiter C, Maglione M, Komlódi T, Ulmer H, Cardini B, Troppmair J, Öfner D, Gnaiger E, Schneeberger S, Oberhuber R (2024) The predictive value of graft viability and bioenergetics testing towards the outcome in liver transplantation.  [[Meszaros_2024_Transpl_Int|»Bioblast link«]]
:from the O2k-Network »[[AT Innsbruck Schneeberger S]] and [[AT Innsbruck Oroboros]]«
|}


May 22nd


:::: [[File:MitoPedia.jpg|60px|MitoPedia]] Why do we fight gas bubbles? - Dive deep into '''[[Oxygen solubility]]'''
:{|
| width="100" align="center" |[[Image:O2k-Publications.jpg|70px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]
|'''"We have identified a circadian clock phase-dependent change in Complex I coupling efficiency, demonstrating for the first time a mitochondria–clock interaction in the hooded seal."''' <br>
* Ciccone C, Kante F, Folkow LP, Hazlerigg DG, West AC, Wood SH (2024) Circadian coupling of mitochondria in a deep-diving mammal. [[Ciccone 2024 J Exp Biol|»Bioblast link«]]
:from the O2k-Network »[[NO Tromsoe Wood SH]]«
|}


== MitoPedia alert ==
== MiPNet ==
:::* Example:
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' "No bubbles should be trapped in the electrolyte reservoir after membrane application and mounting"'''<br>


== O2k-Publications alert ==
- »'''[[MiPNet28.10 SmartPOS-service]]'''«<br>
- communicated by »[[Timon-Gomez Alba]]« and »[[]]«
|}


=== Next ===
=== Next ===


::::[[Image:O2k-Publications.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Publications:_Topics |O2k-Publications in the MiPMap]]''' Mitochondria - a driving force in adipocyte differentiation'''
May 2nd


::::::: - from the O2k-Network »[[CZ Pilsen Kuncova J]]«
:{|
::::::* Kladnická I, Čedíková M, Kripnerová M, Dvořáková J, Kohoutová M, Tůma Z, Müllerová D, Kuncová J (2019) Mitochondrial respiration of adipocytes differentiating from human mesenchymal stem cells derived from adipose tissue. Physiol Res 68:287-96. [[Kladnicka 2019 Physiol Res|»Bioblast link«]]
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' "In respirometric studies on blood cells, the purity of cell preparation should always be emphasized and documented." '''<br>
- »'''[[MiPNet21.17 BloodCellsIsolation]]'''«<br>
- communicated by »[[Timon-Gomez Alba]]« and »[[Baglivo Eleonora]]«
|}


==  O2k-Open Support alert ==
==  O2k-Open Support alert ==
:::*Examples:
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|25px|Info: take care]]
|'''Does your polarographic oxygen sensor (POS) need a deep clean?'''<br>
A POS with a very dark anode might need several rounds of ammonia cleaning, possibly including prolonged exposure overnight to restore performance specifications.
<br>
- »'''[[MiPNet19.18B POS-service|POS service]]'''«<br>
- communicated by »[[Cecatto Cristiane]]« and »[[Leo Elettra]]«
|}
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]
|'''Did you know that a humid environment and/or liquid spillage may lead to a high raw POS signal [V]?'''<br>
See how to clean and dry the electrical connections of the OroboPOS-Connector.<br>
- »'''[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]'''«<br>
- communicated by »[[Cecatto Cristiane]]« and »[[Baglivo Eleonora]]«
|}
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' Unusual stirrer bar rotation? '''<br>
A “jumping” stirrer or a slow rotation of the stirrer bar may indicate a problem either with the stirrer itself or with the cleaning of the glass chamber.
<br>
- »'''[https://wiki.oroboros.at/index.php/Stirrer-Bar%5Cwhite_PVDF%5C15x6_mm#Technical_support Stirrer bar - technical support]'''«<br>
- communicated by »[[Baglivo Eleonora]]« and »[[Grings Mateus]]«
|}


=== Next ===
=== Next ===


::::* While in Home Office: [[User:Gnaigere|Gnaigere]] ([[User talk:Gnaigere|talk]]) 14:03, 19 March 2020 (CET) https://wiki.oroboros.at/index.php/Talk:O2k-Open_Support_alert  -  Many labs will move to home office mode. In line with our special DL7.4 home office offer (see Agenda today), therefore, I suggest to post in our Agenda '''analysis-linked''' O2k-Open Support cases during this time, and collect real-time operation cases for future releases.


:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' Do you observe an unstable O<sub>2</sub> slope negative following TIP2k titrations during an instrumental O<sub>2</sub> background test?  '''<br>
Incorrect placement of TIP2k Filter Papers can lead to liquid accumulation and convective oxygen backdiffusion. Check for proper placement by confirming they are fully in contact with the stopper receptacle.
<br>
- »'''[[MiPNet14.06_Instrumental_O2_background|Instrumental O<sub>2</sub> background]]'''«, »'''[https://wiki.oroboros.at/index.php/Oxygen_flux_-_instrumental_background#Unstable_O2_slope_neg._in_instrumental_background_test Unstable O<sub>2</sub> slope negative]'''«<br>
- communicated by »[[Grings Mateus]]« and »[[Baglivo Eleonora]]«
|}




* '''May-18 Mo'''
April 15th
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|'''Do you observe either a zero or low raw O<sub>2</sub> signal when starting a measurement?'''<br>
Perform a test to evaluate whether this is related to an improperly positioned chamber or a faulty connection between the OroboPOS and the OroboPOS-Connector. <br>
- »'''[https://wiki.oroboros.at/index.php/Oxygen_signal#Raw_signal Oxygen signal]'''«<br>
- communicated by »[[Grings Mateus]]« and »[[Schmitt Sabine]]«
|}


:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you know that a humid O2k-environment and/or liquid spillage may lead to a high raw POS signal [V]?'''
April 29th
::::::: See how to clean and dry the electrical connections of the OroboPOS-Connector.-»[[OroboPOS-Connector#Cleaning_the_electrical_connections|Cleaning the electrical connections]]«
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
=== Already added to [https://wiki.oroboros.at/index.php?title=Template:Next_oroboros_ecosystem_agenda Agenda - Next] and to be added to [https://wiki.oroboros.at/index.php/O2k-Open_Support_alert O2k-Open support alert] ===
|'''How can I update DatLab 8 to the latest version?'''<br>
* '''May-04 Mo'''
If your computer with DatLab 8 is connected to the internet, go to “Check for updates” in the Help menu of DatLab 8, and follow the simple installation instructions. <br>
- »'''[[Updating - DatLab]]'''«, »'''[https://www.oroboros.at/index.php/product/datlab/ DatLab 8]'''«<br>
- communicated by »[[Leo Elettra]]« and »[[Baglivo Eleonora]]«
|}


:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|40px|Take care]] '''Do you load the correct instrumental background correction accordingly to your range of experimental O<sub>2</sub> levels?'''
== O2k-Feedback alert ==
::::::: See how to perform the instrumental background test. - »[[Flux / Slope| O2 background correction in DatLab]]« and »[[Oxygen_flux_-_instrumental_background | Instrumental oxygen background]]«
:::* Example (if the person has a profile, link the name to the profile with [[Cecatto C| Cristiane Cecatto]]:
:{|
| width="110" align="center" | [[Image:O2k-Feedback.jpg|60px|link=https://wiki.oroboros.at/index.php/O2k-Feedback |O2k-Feedback]]
|''' ''“Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”'''''<br>
Badr Ibrahim, PhD from Betterhumans Inc., US
<br>
- »'''[[O2k-Feedback]]'''«<br>
|}
:{|
| width="110" align="center" | [[Image:O2k-Feedback.jpg|60px|link=https://wiki.oroboros.at/index.php/O2k-Feedback |O2k-Feedback]]
|''' ''“Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”'''''<br>
Badr Ibrahim, PhD from the O2k-Network »[[UK Leicester Moisoi N]]«<br>
- »'''[[O2k-Feedback]]'''«<br>
|}


* '''May-11 Mo'''
=== Next ===
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you know how the "switch components" approach can help you find what is causing your O2k to not reach the quality control specifications?'''
::::::: Do not despair if specifications given in the O2k-Manuals are not obtained. See how to troubleshoot on your own.-»[[Oxygen_sensor_test#Switch_components_approach|Switch components approach]]«
 
=== Removed for correction ===
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform the ''leaky chamber test'' after chamber assembly? '''
::::::: - An unnoticed leak from the chamber may affect your data, therefore we recommend performing the ''Leaky chamber test'' - especially after chamber reassembly. - »[[Talk:O2k-Chamber_MitoPedia#Leaky_Chamber|Leaky chamber test]]«
<br />
:::::::» https://intern.oroboros.at/index.php/Overnight_test_after_O2k-chamber_assembly
 
== O2k-Open Support alert circulars ==
 
=== O2k-Open Support alert circular 2020(01): in prep ===
::'''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.'''
::::::: See how you can check that the two blue OroboPOS-Holders are intact. - »[[O2k-OroboPOS-Holder_MitoPedia#O2k-OroboPOS-Holder:_Leaky_chamber_due_to_damaged_OroboPOS-Holder |O2k-OroboPOS-Holder]]«
<br />
::'''''New''''' [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Yellow-take care.png|40px|Take care]] '''Calibration parameters are transferred to experimental DatLab files'''
::::::: Save calibration information from selected [[Marks - DatLab |Marks]] while DatLab is connected to the O2k during an instrumental calibration run. Air calibration and zero oxygen calibration can be performed and saved in a single DatLab file or in separate DatLab files. - »[[Oxygen_calibration_-_DatLab#Oxygen_calibration_in_instrumental_and_experimental_DatLab_files | Oxygen calibration - DatLab]]«
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform O2k-Chamber volume calibration?'''
::::::: The volume calibration is essential to ensure that the chamber volume is correctly calibrated to its standard volume of 2 mL (or 0.5 mL for the [[O2k-sV-Module]]). - »[[O2k-Chamber MitoPedia]]«
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you use the correct O<sub>2</sub> solubility factor (F<sub>M</sub>) in the ‘O2 calibration’ window in DatLab?
 
::::::: The O<sub>2</sub> solubility factor varies according to the medium used and influences the air calibration. - »[[Oxygen solubility factor]]«== In preparation ==
 
== For discussion ==
''Here we discuss suggestions for the Oroboros Ecosystem Agenda 2020''
 
::: Controlled stability of the [https://www.oroboros.at/index.php/product/mitokit-cii MitoKit-CII] based on HPLC and LCMS analysis: No degradation of [[MitoKit-CII_MitoPedia| MitoKit-CII]] compounds [[MitoKit-CII/Succinate-nv| MitoKit-CII/Succinate-nv (NV118)]] and [[MitoKit-CII/Malonate-nv| MitoKit-CII/Malonate-nv (NV161)]] for following storage conditions:
:::* Proven -80 °C solid stability for more than 3 years (3 years 7 months)
:::* Proven -20 °C solid stability for 1 year
:::* Proven DMSO stability at room temperature for 1 week
:::* Proven DMSO stability at -20 °C for 6 months (ongoing)


== MitoFit Preprints ==
:::* Example:
:{|
| width="100" align="center"|[[Image:BEC-logo.png|100px|link=https://www.bioenergetics-communications.org/index.php/bec |Bioenergetics Communications]]
|'''From preprint to BEC manuscript with published Open Peer Reviews:'''<br>
*Komlódi T, Sobotka O, Gnaiger E (2021) Facts and artefacts on the oxygen dependence of hydrogen peroxide production using Amplex UltraRed. Bioenerg Commun 2021.4. [[doi:10.26124/BEC:2021-0004]]
:from the O2k-Network »[[AT Innsbruck Oroboros]]«
|}


=== Next ===


== Under discussion ==


:::: [[File:MitoPedia.jpg|left|60px|MitoPedia]] As part of the [[NextGen-O2k|H2020 SME NextGen-O2k]] project we are developing a PhotoBiology module that will allow to study the effect of different light intensities on O<sub>2</sub> consumption and production: »'''[[PhotoBiology]]'''«
This one needs an update of the Bioblast page and clarification of MiPNet06.03


:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Yellow-take care.png|40px|Take care]] '''Do you perform a stirrer test during oxygen calibration?'''
:{|
::::::: In DatLab 7.4, instrumental DL-Protocols guide you automatically towards performing the stirrer test in the process of air calibration. It shows immediately, if your OroboPOS has a satisfactory response time. ''Further details in'' [[MitoPedia: DatLab |MitoPedia]]: »[[Oxygen calibration - DatLab]]
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' Did you enter the correct oxygen solubility factor in DatLab?'''<br>
The oxygen solubility factor of the incubation medium, F<sub>M</sub>, expresses the effect of the salt concentration on oxygen solubility relative to pure water. The correct oxygen solubility factor of your experimental medium or buffer needs to be entered in DatLab for proper calculation of the oxygen concentration.
<br>
- »'''[[Oxygen solubility factor|Oxygen solubility factor]]'''«<br>
- communicated by »[[Schmitt Sabine]]« and »[[ ]]«


:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Is the noise of the ‘O2 slope neg’ acceptable?'''
|}
::::::: If you are not certain that the noise of the ‘O2 slope neg’ is within a suitable range or you want to confirm that your experimental data will not be affected, see ''further details'': [[MiPNet06.03 POS-calibration-SOP]]


== 2021 ==
This one needs a BioBlast link:


::::* 2021-01-06 https://www.jstor.org/stable/1706346?seq=1
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|''' When filling in the volume of uncoupler in the DatLab mark or the excel template, should one enter the total volume added or the volume to get to the highest flux?'''<br>
The volume added until the mark, equal to the highest flux, should be entered. Anything titrated after this mark should be summed to the other volumes before the next mark because this will be considered to calculate the sample dilution. 
<br>
- »'''[[LINK|LINK]]''' and '''[[MiPNet24.06_Oxygen_flux_analysis_-_DatLab_7.4|O<sub>2</sub> flux analysis with DatLab 7.4]]'''«<br>
- communicated by »[[Cardoso Luiza]]« and »[[ ]]«
|}


== Postponed events ==


:::: [[File:FAT4BRAIN logo.png|70px|FAT4BRAIN|link=FAT4BRAIN School Riga LV]] »<big>'''[[FAT4BRAIN School Riga LV | FAT4BRAIN School Riga LV IOC147]]'''</big
== Removed for correction ==
* '''Nov-30 Mo'''
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Blue-info.png|25px|Info: improve]]
|'''Noisy oxygen signal? Check the OroboPOS and OroboPOS-Connector to clean electrical connections.'''<br>
Contaminating grease, salt crystals and moisture can be removed with a dry paper cloth or a paper cloth moistened with water and then absolute ethanol. <br>
- »[[OroboPOS-Connector#Cleaning_the_electrical_connections |Cleaning the electrical connections]]« <br>
- communicated by »[[Schmitt Sabine]]« and »[[Komlodi Timea]]«
|}
WGT recommended not to use water for cleaning. Erich introduced water for cleaning to remove salt crystals.


:::: »<big>'''[[APS2020 Chicago US]]'''</big>«


:::: [[Image:O2k-Network.png|70px|O2k-Workshops|link=O2k-Events#Next_O2k-Workshops]] [[MiPNet25.09 IOC San Diego CA US |<big>'''San Diego CA US'''</big>]]
:{|
| width="110" align="center" | [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]][[File:Red-loss of data.png|25px|Warning: loss of data]]
|'''Do you perform the ''leaky chamber test'' after chamber assembly? '''<br>
An unnoticed leak from the chamber may affect your data, therefore we recommend performing the ''Leaky chamber test'' - especially after chamber reassembly. <br>
- »[[Talk:O2k-Chamber_MitoPedia#Leaky_Chamber|Leaky chamber test]]« and » https://intern.oroboros.at/index.php/Overnight_test_after_O2k-chamber_assembly«
|}

Latest revision as of 17:24, 23 April 2024

Antunes Diana (talk) 08:58, 5 May 2020 (CEST) Format for new Agenda/Alerts entries:
 {|
 | logos
 | title in bold <br>
 - link
 |}
 

e.g.,

O2k-Open SupportWarning: loss of data Did you know that the proper use of TIP2k-Filter Papers may avoid an unstable O2 slope neg. signal during Instrumental background correction?

- »Instrumental background« and »TIP2k Filter Papers«

See the scheme:


Agenda/Alerts body text format


MitoPedia alert

Example:
MitoPedia
Measuring fatty acid oxidation (FAO)-linked respiration requires a low concentration of malate

In the absence of malate, accumulation of Acetyl-CoA inhibits FAO.
- »Fatty acid oxidation pathway control state«, »Mitochondrial Pathways: Chapter 5« 
- communicated by »Cecatto Cristiane« and »Cardoso Luiza« 


Next

Under construction

on hold until the page is updated
FAT4BRAIN
A FAO-mito-CNS platform to help fatty acid-related respiration measurements was developed in the FAT4BRAIN project

- » FAT4BRAIN FAO-mito-CNS platform «

O2k-Brief alert

  • Example:
O2k-brief repository Antitumoral properties of triterpenic acid functionalized gold nanoparticles arise from decreasing mitochondrial respiration
  • Mioc M, Mioc A, Racoviceanu R, Ghiulai R, Prodea A, Milan A, Barbu Tudoran L, Oprean C, Ivan V, Șoica C (2023) The antimelanoma biological assessment of triterpenic acid functionalized gold nanoparticles. »Bioblast link«

Next

O2k-Publications alert

  • Example:
O2k-Publications in the MiPMap "Mitochondria–nucleus communication and its integration within the maintenance of cellular homeostasis constitute key determinants of cellular fate and organism health in the context of cellular stress, homeostasis perturbation, and disease"
  • Temelie M, Talpur R, Dominguez-Prieto M, Dantas Silva A, Cenusa C, Craciun L, Savu DI, Moisoi N (2023) Impaired integrated stress response and mitochondrial integrity modulate genotoxic stress impact and lower the threshold for immune signalling. »Bioblast link«
from the O2k-Network »UK Leicester Moisoi N«

Author names of O2k-Publications/ Brief for the Agenda vs Bioblast

Click to expand or collaps

Guidelines to update a reference

When creating an O2k-Publication/Brief alert, it is the responsibility of the author to check if the reference is accordingly to the correct guidelines:

Reference example.png

Correct format for a person called “First Name Surname”:
In the gray box should be “Surname FN” - From where references to be used as O2k-Publication/Brief alerts are copied from.

Gray box reference.png

In the links should be “ Surname First N”

Authors names links reference.png

How to edit:
A) Check if all authors with profiles in Bioblast (blue links) have their profiles with full name. If they have, proceed to next step (B). If not, you need to move it following the steps “How to move a profile” below.
B) Click in “Edit with form”.
C) The gray box is edited in “Reference” and the links are changed in “Author(s)”. Change it accordingly and click in “save page” at the end of the page.

Editing reference.png

How to move a profile (instructions sent by Marija):

So let’s take for example my page: Beno M [1]

As you can see there is a redirect from Beno Marija to Beno M. What has to be done is to switch the content of those pages. Here are the steps:
1. Go to e.g. “Beno M” on edit
2. Click on “Move”
3. Change the new title to the full name of the person. In my case it will be Beno M to Beno Marija

BenoMreference.jpg

4. Now it should look like that:

BenoMarijaReference.png

5. Click on “Move” and leave all the ticked boxes
6.Done
After that you can continue in step B.
If you have any doubt you can contact Cristiane, Marija or Mario.

Next

April 25th

O2k-Publications in the MiPMap "Our results further enhance the understanding and relevance of bioenergetic function in liver ischemia and transplantation and provide the basis for further consideration of these parameters as biomarkers in LT [liver transplantation]."
  • Meszaros AT, Weissenbacher A, Schartner M, Egelseer-Bruendl T, Hermann M, Unterweger J, Mittelberger C, Reyer BA, Hofmann J, Zelger BG, Hautz T, Resch T, Margreiter C, Maglione M, Komlódi T, Ulmer H, Cardini B, Troppmair J, Öfner D, Gnaiger E, Schneeberger S, Oberhuber R (2024) The predictive value of graft viability and bioenergetics testing towards the outcome in liver transplantation. »Bioblast link«
from the O2k-Network »AT Innsbruck Schneeberger S and AT Innsbruck Oroboros«

May 22nd

O2k-Publications in the MiPMap "We have identified a circadian clock phase-dependent change in Complex I coupling efficiency, demonstrating for the first time a mitochondria–clock interaction in the hooded seal."
  • Ciccone C, Kante F, Folkow LP, Hazlerigg DG, West AC, Wood SH (2024) Circadian coupling of mitochondria in a deep-diving mammal. »Bioblast link«
from the O2k-Network »NO Tromsoe Wood SH«

MiPNet

  • Example:
O2k-Open SupportInfo: improve "No bubbles should be trapped in the electrolyte reservoir after membrane application and mounting"

- »MiPNet28.10 SmartPOS-service«
- communicated by »Timon-Gomez Alba« and »[[]]«

Next

May 2nd

O2k-Open SupportInfo: improve "In respirometric studies on blood cells, the purity of cell preparation should always be emphasized and documented."

- »MiPNet21.17 BloodCellsIsolation«
- communicated by »Timon-Gomez Alba« and »Baglivo Eleonora«

O2k-Open Support alert

  • Examples:
O2k-Open SupportInfo: take care Does your polarographic oxygen sensor (POS) need a deep clean?

A POS with a very dark anode might need several rounds of ammonia cleaning, possibly including prolonged exposure overnight to restore performance specifications.
- »POS service«
- communicated by »Cecatto Cristiane« and »Leo Elettra«

O2k-Open SupportWarning: loss of data Did you know that a humid environment and/or liquid spillage may lead to a high raw POS signal [V]?

See how to clean and dry the electrical connections of the OroboPOS-Connector.
- »Cleaning the electrical connections«
- communicated by »Cecatto Cristiane« and »Baglivo Eleonora«

O2k-Open SupportInfo: improve Unusual stirrer bar rotation?

A “jumping” stirrer or a slow rotation of the stirrer bar may indicate a problem either with the stirrer itself or with the cleaning of the glass chamber.
- »Stirrer bar - technical support«
- communicated by »Baglivo Eleonora« and »Grings Mateus«

Next

O2k-Open SupportInfo: improve Do you observe an unstable O2 slope negative following TIP2k titrations during an instrumental O2 background test?

Incorrect placement of TIP2k Filter Papers can lead to liquid accumulation and convective oxygen backdiffusion. Check for proper placement by confirming they are fully in contact with the stopper receptacle.
- »Instrumental O2 background«, »Unstable O2 slope negative«
- communicated by »Grings Mateus« and »Baglivo Eleonora«


April 15th

O2k-Open SupportInfo: improve Do you observe either a zero or low raw O2 signal when starting a measurement?

Perform a test to evaluate whether this is related to an improperly positioned chamber or a faulty connection between the OroboPOS and the OroboPOS-Connector.
- »Oxygen signal«
- communicated by »Grings Mateus« and »Schmitt Sabine«

April 29th

O2k-Open SupportInfo: improve How can I update DatLab 8 to the latest version?

If your computer with DatLab 8 is connected to the internet, go to “Check for updates” in the Help menu of DatLab 8, and follow the simple installation instructions.
- »Updating - DatLab«, »DatLab 8«
- communicated by »Leo Elettra« and »Baglivo Eleonora«

O2k-Feedback alert

  • Example (if the person has a profile, link the name to the profile with Cristiane Cecatto:
O2k-Feedback “Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”

Badr Ibrahim, PhD from Betterhumans Inc., US
- »O2k-Feedback«

O2k-Feedback “Thank you very much for your thorough help. I appreciate the time you took to provide me with all the answers and supportive literature to realize the potential of the O2k-respirometer in bioenergetics research”

Badr Ibrahim, PhD from the O2k-Network »UK Leicester Moisoi N«
- »O2k-Feedback«

Next

MitoFit Preprints

  • Example:
Bioenergetics Communications From preprint to BEC manuscript with published Open Peer Reviews:
  • Komlódi T, Sobotka O, Gnaiger E (2021) Facts and artefacts on the oxygen dependence of hydrogen peroxide production using Amplex UltraRed. Bioenerg Commun 2021.4. doi:10.26124/BEC:2021-0004
from the O2k-Network »AT Innsbruck Oroboros«

Next

Under discussion

This one needs an update of the Bioblast page and clarification of MiPNet06.03

O2k-Open SupportInfo: improve Did you enter the correct oxygen solubility factor in DatLab?

The oxygen solubility factor of the incubation medium, FM, expresses the effect of the salt concentration on oxygen solubility relative to pure water. The correct oxygen solubility factor of your experimental medium or buffer needs to be entered in DatLab for proper calculation of the oxygen concentration.
- »Oxygen solubility factor«
- communicated by »Schmitt Sabine« and »[[ ]]«

This one needs a BioBlast link:

O2k-Open SupportInfo: improve When filling in the volume of uncoupler in the DatLab mark or the excel template, should one enter the total volume added or the volume to get to the highest flux?

The volume added until the mark, equal to the highest flux, should be entered. Anything titrated after this mark should be summed to the other volumes before the next mark because this will be considered to calculate the sample dilution.
- »LINK and O2 flux analysis with DatLab 7.4«
- communicated by »Cardoso Luiza« and »[[ ]]«


Removed for correction

  • Nov-30 Mo
O2k-Open SupportInfo: improve Noisy oxygen signal? Check the OroboPOS and OroboPOS-Connector to clean electrical connections.

Contaminating grease, salt crystals and moisture can be removed with a dry paper cloth or a paper cloth moistened with water and then absolute ethanol.
- »Cleaning the electrical connections« 
- communicated by »Schmitt Sabine« and »Komlodi Timea«

WGT recommended not to use water for cleaning. Erich introduced water for cleaning to remove salt crystals.


O2k-Open SupportWarning: loss of data Do you perform the leaky chamber test after chamber assembly?

An unnoticed leak from the chamber may affect your data, therefore we recommend performing the Leaky chamber test - especially after chamber reassembly.
- »Leaky chamber test« and » https://intern.oroboros.at/index.php/Overnight_test_after_O2k-chamber_assembly«