Talk:O2k-Open Support alert: Difference between revisions

From Bioblast
No edit summary
No edit summary
Line 1: Line 1:
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform the ''leaky chamber test'' after chamber assembly? '''
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Red-loss of data.png|40px|Warning: loss of data]] '''Do you perform the ''leaky chamber test'' after chamber assembly? '''
::::::: After chamber reassembly you might experience a small leaky chamber only identified by a ''leaky chamber test''- ยป[https://www.bioblast.at/index.php/O2k-Chamber_MitoPedia#Leaky_chamber_test Leaky chamber test]ยซ
::::::: After chamber reassembly you might experience a small leaky chamber only identified by a ''leaky chamber test''- ยป[[O2k-Chamber_MitoPedia#Leaky_chamber_test|Leaky chamber test]]ยซ


:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Blue-info.png|40px|Info: improve]] '''Do you change the [[Smoothing|slope smoothing]] to 20 while running experiments with biological sample?'''
:::: [[Image:O2k-Support.jpg|60px|link=http://wiki.oroboros.at/index.php/O2k-Open_Support |O2k-Open Support]] [[File:Blue-info.png|40px|Info: improve]] '''Do you change the [[Smoothing|slope smoothing]] to 20 while running experiments with biological sample?'''

Revision as of 10:48, 5 March 2020

O2k-Open Support Warning: loss of data Do you perform the leaky chamber test after chamber assembly?
After chamber reassembly you might experience a small leaky chamber only identified by a leaky chamber test- ยปLeaky chamber testยซ
O2k-Open Support Info: improve Do you change the slope smoothing to 20 while running experiments with biological sample?
While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - ยปSlope smoothingยซ
O2k-Open Support Take care Did you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?
By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with specific settings for the O2k-sV-Module. - ยปO2k-sV-Moduleยซ