Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Oxygen flux - instrumental background"

From Bioblast
Line 1: Line 1:
{{MitoPedia
{{MitoPedia
|abbr=''J''°<sub>O2</sub>
|abbr=''J''°<sub>O2</sub>
|description='''Instrumental background oxygen flux''', ''J''°<sub>O2</sub>, in a respirometer is due to oxygen consumption by the [[POS]], and oxygen diffusion into or out of the aqueous medium in the [[O2k-chamber]].  It is measured in the range of experimental oxygen levels by a standardized [[instrumental background test]], and is a property of the instrumental system.  Instrumental background correction eliminates errors by systemic flux compensation, automatically performed by [[DatLab]]. See also [[Chemical background correction of oxygen flux]].
|description='''Instrumental background oxygen flux''', ''J''°<sub>O2</sub>, in a respirometer is due to oxygen consumption by the [[POS]], and oxygen diffusion into or out of the aqueous medium in the [[O2k-Chamber]].  It is measured in the range of experimental oxygen levels by a standardized [[instrumental background test]], and is a property of the instrumental system.  Instrumental background correction eliminates errors by systemic flux compensation, automatically performed by [[DatLab]]. See also [[Chemical background correction of oxygen flux]].
|info=[[MiPNet14.06]]; [[Gnaiger_2008_POS]]; [[Gnaiger_2001_RespPhysiol]]
|info=[[MiPNet14.06]]; [[Gnaiger_2008_POS]]; [[Gnaiger_2001_RespPhysiol]]
|type=Respiration
|type=Respiration

Revision as of 16:21, 7 January 2012


high-resolution terminology - matching measurements at high-resolution


Oxygen flux - instrumental background

Description

Instrumental background oxygen flux, J°O2, in a respirometer is due to oxygen consumption by the POS, and oxygen diffusion into or out of the aqueous medium in the O2k-Chamber. It is measured in the range of experimental oxygen levels by a standardized instrumental background test, and is a property of the instrumental system. Instrumental background correction eliminates errors by systemic flux compensation, automatically performed by DatLab. See also Chemical background correction of oxygen flux.

Abbreviation: J°O2

Reference: MiPNet14.06; Gnaiger_2008_POS; Gnaiger_2001_RespPhysiol


MitoPedia methods: Respirometry 



Instrumental background oxygen flux,Instrumental background oxygen flux, J°<sub>O2</sub>, in a respirometer is due to oxygen consumption by the POS, and oxygen diffusion into or out of the aqueous medium in the O2k-Chamber. It is measured in the range of experimental oxygen levels by a standardized instrumental background test, and is a property of the instrumental system. Instrumental background correction eliminates errors by systemic flux compensation, automatically performed by DatLab. Automatic correction for the instrumental background oxygen flux is an essential standard in high resolution respirometry. At the same time an instrumental background experiment is the ultimate test for instrumental performance, evaluating chamber performance after completion of all elements of the Sensor test.[Sensor test]]. +
J°O2 +
Respirometry +
O2 signal +  and Performance test +