|description=In '''O2k configuration''', channels (amperometric and potentiometric) can be switched on/off by selecting the according tick box.
|description=In '''O2k configuration''', channels (amperometric and potentiometric) can be switched on/off by selecting the according tick box.


Power-O2k and numbers for O2 sensor, Amp sensor, pX electrode and pX reference electrode are entered and edited here. In case the O2k-Series H with the fluorescence sensors is used, the fluorescence sensor serial numbers are shown automatically under "Amperometric, Amp". The O2k configuration window pops up when DatLab starts and "Connect to O2k" is pressed for the first time. It is also accessible from the menu "Oroboros O2k" and from within the [[O2k control]] and [[Mark statistics - DatLab|Mark statistics]] windows.
Power-O2k and numbers for O2 sensor, Amp sensor, pX electrode and pX reference electrode are entered and edited here. With the [[O2k-FluoRespirometer]] (O2k-Series H), the fluorescence sensor serial numbers are shown automatically under "Amperometric, Amp". The O2k configuration window pops up when DatLab starts and "Connect to O2k" is pressed for the first time. It is also accessible from the menu "Oroboros O2k" and from within the [[O2k control]] and [[Mark statistics - DatLab|Mark statistics]] windows.
|info=[[MiPNet19.18C DatLab-guide]]
|info=[[http://wiki.oroboros.at/index.php/MitoPedia:_DatLab#Oroboros_O2k |Oroboros O2k]]
}}
}}
{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
Line 12: Line 12:
ย  Communicated by [[Capek O |Ondrej Capek]] 2017-07-11.
ย  Communicated by [[Capek O |Ondrej Capek]] 2017-07-11.
ย ย  ย 
ย ย  ย 
:::: Problem (special case): In case an O2k-Series H was switched on with fluorescence sensors already connected, DatLab started afterwards and "Connect to O2k" clicked, the corresponding serial numbers of the fluorescence sensors under "Amperometric, Amp" are not yet shown. ย 
:::: Problem (special case): (i) O2k-Series H Main-Unit is switched on with fluorescence sensors already connected; (ii) DatLab is started afterwards and "Connect to O2k" clicked; (iii) ''Problem:'' DatLab does not show the corresponding serial numbers of the fluorescence sensors under "Amperometric, Amp". ย 
:::: Solution: After confirming the "O2k configuration" window with "OK", the [[O2k control]] window opens automatically - another click on "Configuration" opens the "O2k configuration" windows again, with the correct fluorescence sensors serial numbers.
:::: Solution: (i) Open the "O2k configuration" window (this can be done directly in the [[O2k control]] window by a click on "O2k configuration"; (ii) now the correct fluorescence sensor serial numbers are shown automatically.

Revision as of 11:09, 14 July 2017


high-resolution terminology - matching measurements at high-resolution


O2k configuration

Description

In O2k configuration, channels (amperometric and potentiometric) can be switched on/off by selecting the according tick box.

Power-O2k and numbers for O2 sensor, Amp sensor, pX electrode and pX reference electrode are entered and edited here. With the O2k-FluoRespirometer (O2k-Series H), the fluorescence sensor serial numbers are shown automatically under "Amperometric, Amp". The O2k configuration window pops up when DatLab starts and "Connect to O2k" is pressed for the first time. It is also accessible from the menu "Oroboros O2k" and from within the O2k control and Mark statistics windows.


Reference: [|Oroboros O2k]


MitoPedia O2k and high-resolution respirometry: DatLab 


O2k-Series H: Fluorescence sensors serial numbers

Communicated by Ondrej Capek 2017-07-11.
 
Problem (special case): (i) O2k-Series H Main-Unit is switched on with fluorescence sensors already connected; (ii) DatLab is started afterwards and "Connect to O2k" clicked; (iii) Problem: DatLab does not show the corresponding serial numbers of the fluorescence sensors under "Amperometric, Amp".
Solution: (i) Open the "O2k configuration" window (this can be done directly in the O2k control window by a click on "O2k configuration"; (ii) now the correct fluorescence sensor serial numbers are shown automatically.