May 2020

  • May-11 Th
O2k-Open SupportTake care Do you know how to avoid underestimating/overestimating your respiratory fluxes? Measuring steady states with high-resolution respirometry is the key to obtaining correct values.

- Learn more about steady states measured with high-resolution respirometry: performance and mark setting for data analysis. - Β»Steady stateΒ«


  • May-04 Mo
O2k-Open SupportTake care Do you load the correct instrumental background correction accordingly to your range of experimental O2 levels?

- See how to perform the instrumental background test. - Β» O2 background correction in DatLabΒ« and Β» Instrumental oxygen backgroundΒ«

Apr 2020

  • Apr-27 Mo
O2k-Open SupportTake care Do you select the appropriate marks for the instrumental background correction when planning to apply them to a protocol with a high O2 regime?

- To obtain an accurate O2 background correction, you must select the marks that correspond to your range of experimental O2 levels. - Β» O2 Background correction - troubleshootingΒ«


  • Apr-20 Mo
O2k-Open SupportImprove Did you know that some substrates alone cannot maintain a stable mitochondrial respiratory flux in OXPHOS?

- Unstable respiratory fluxes in OXPHOS might be due to substrates limitation. - Β» Pyruvate - troubleshootingΒ«


  • Apr-14 Mo
O2k-Open SupportTake care Why are the oxygen levels different between chambers prior to my experimental run?

- After switching on the O2k, air calibration has to be performed; otherwise the displayed concentrations may deviate from accurate concentrations. - Β» Oxygen calibration - troubleshootingΒ«


  • Apr-07 Mo
DatLabImprove Did you know that you can edit the background parameters in the 'O2 analysis template DL7.4?

- If you did not apply a finally evaluated O2 background correction to your experimental DatLab file, you can edit the O2 background parameters directly in the O2 analysis template DL7.4 for completion of Oxygen flux analysis - DatLab 7.4. - Β»SUIT: Browse DL-Protocols and templatesΒ«


  • Apr-02 Mo
MitoPedia: SUITImprove Did you know that you can edit the background parameters in the 'O2 analysis template DL7.4?

- Do you want to search the best SUIT protocol for your research questions? - Use the SUITbrowser: Β»SUITbrowserΒ«

Mar 2020

  • Mar-30 Mo
O2k-Open SupportInfo: improve DatLab 7.4 offers the flexibility to edit a DL-Protocol:

- Β»Enable DL-Protocol editingΒ«


  • Mar-25 We
O2k-Open SupportInfo: improve O2 flux analysis with DatLab 7.4. Use our latest O2 analysis template DL7.4:

- Β»O2k-Videosupport: O2 flux analysis with DatLab 7.4Β«


  • Mar-23 Mo
O2k-Open SupportTake care Do you know that you can use the Titration-Injection microPump (TIP2k) to perform an instrumental O2 background on the O2k-sV-Module?

- (1) By adjusting the dithionite concentration to 2.5 mM, the instrumental O2 background can be performed with the "BG Feedback" settings. - Β»O2k-sV-ModuleΒ« 
- (2) By adjusting the TIP2k setup with specific settings for the O2k-sV-Module, the instrumental O2 background can be performed in the O2k-sV-Module. - Β»O2k-sV-ModuleΒ«


  • Mar-11 We
O2k-Open SupportInfo: improve Do you change the slope smoothing to 20 while running experiments with biological sample?

- While running your experiment, especially during uncoupler titrations, it is important to decrease the slope smoothing in order to have a faster visualization of the titrations effect on respiration. - Β»Slope smoothingΒ«




  • Mar-04 We
OroboPOS-Holder002.jpg


New O2k-Open Support Warning: loss of data A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber.
See how you can check that the two blue OroboPOS-Holders are intact. - Β»O2k-OroboPOS-HolderΒ«


Feb 2020

  • Feb-26 We
MitoPedia It is essential to titrate uncouplers in multiple steps for respirometric measurement of electron transfer capacity.

- Β»Noncoupled respirationΒ«



  • Feb-24 Mo
MitoPedia More research is needed on the impact of viral infection on mitochondrial function.

- Β»Viruses and mitochondrial medicineΒ«



  • Feb-19 We
MitoPedia Spermidine and other calorie restriction mimetics control autophagy and mitochondrial function.

- Β»SpermidineΒ«



  • Feb-18 Tu
MitoPedia Why do we fight gas bubbles?

- Dive deep into Β»Oxygen solubilityΒ«



  • Feb-12 We
MitoPedia What do metformin, melatonin and other bioactive mitObesity compounds have in common?

- Β»MitObesity drugsΒ«


Jan 2020

  • Jan-16 Th
MitoPedia Oxygen flux

- which SI units should be used?


MitoPedia Did you ever think that units for body mass could be an issue?

- Β»Body massΒ«



  • Jan-08 We
MitoPedia Force and Energy: this we all need in the New Year - and MitoPedia may help to clarify the difference.