== Sensor service by OROBOROS ==
The anode of the POS consists of a solid silver ring coated withΒ  silver chloride (There is no "Ag coating"). The AgCl coating will GROWΒ  during use. A typical contamination will be the formation of AgO. AgOΒ  will be preferentially removed by the ammonia treatment during sensorΒ  service, however part of the AgCl layer will also be removed. AnyΒ  defects in the AgCl layer are restored by running the O2k for a fewΒ  hours (or over night) after the sensor service or just by continued use.
We offer a sensor service in our workshop. This is essentially the same that you can do on your own. The main advantage of the sensor service in our workshop is the presence of a very experienced and very patient team, carrying out many iterations of NH3 cleaning. Running the O2k over night may be required to stabilize a newly serviced OroboPOS.

Revision as of 22:32, 6 February 2016

                



MiPNet19.18B POS-service

Publications in the MiPMap
O2k-Manual
Service of the polarographic oxygen sensor OroboPOS. Β»Bioblast pdfΒ«

Β» Versions

OROBOROS (2015-11-02) Mitochondr Physiol Network

Abstract: Gnaiger E (2015) Service of the polarographic oxygen sensor OroboPOS. Mitochondr Physiol Network 19.18(B03):17-22.

Β» O2k-Videosupport
Β» O2k-SOP: Polarographic oxygen sensors: calibration, accuracy and quality control SOP. Β»POS-Calibration-SOPΒ«
Β» Product: OroboPOS, O2k, O2k-Catalogue

β€’ Keywords: HRR

β€’ O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual 

O2k-Core 



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Sensor service by OROBOROS

The anode of the POS consists of a solid silver ring coated with silver chloride (There is no "Ag coating"). The AgCl coating will GROW during use. A typical contamination will be the formation of AgO. AgO will be preferentially removed by the ammonia treatment during sensor service, however part of the AgCl layer will also be removed. Any defects in the AgCl layer are restored by running the O2k for a few hours (or over night) after the sensor service or just by continued use.

We offer a sensor service in our workshop. This is essentially the same that you can do on your own. The main advantage of the sensor service in our workshop is the presence of a very experienced and very patient team, carrying out many iterations of NH3 cleaning. Running the O2k over night may be required to stabilize a newly serviced OroboPOS.