Talk:Lower O2 limit - DatLab: Difference between revisions
From Bioblast
No edit summary |
(Replaced content with "== DatLab 8 perspectives: == :::: For each respiratory state in a SUIT protocol, the lower O2 limit should be defined in the software according to preparation (imt, cells,...") Tag: Replaced ย |
||
(3 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
== DatLab 8 perspectives: == | |||
ย | |||
== DatLab 8 == | |||
:::: For each respiratory state in a SUIT protocol, the lower O2 limit should be defined in the software according to preparation (imt, cells, permeabilized fibers) and according to the consequent respiratory state. Before uncoupler titration and before strongly stimulating oxygen flux, the lower limit is suggested to be higher than before inhibition of respiration. | :::: For each respiratory state in a SUIT protocol, the lower O2 limit should be defined in the software according to preparation (imt, cells, permeabilized fibers) and according to the consequent respiratory state. Before uncoupler titration and before strongly stimulating oxygen flux, the lower limit is suggested to be higher than before inhibition of respiration. | ||
Latest revision as of 08:14, 7 October 2021
DatLab 8 perspectives:
- For each respiratory state in a SUIT protocol, the lower O2 limit should be defined in the software according to preparation (imt, cells, permeabilized fibers) and according to the consequent respiratory state. Before uncoupler titration and before strongly stimulating oxygen flux, the lower limit is suggested to be higher than before inhibition of respiration.