Divakaruni AS, Jastroch M (2022) A practical guide for the analysis, standardization and interpretation of oxygen consumption measurements. Nat Metab 4:978-94. https://doi.org/10.1038/s42255-022-00619-4

Β» PMID: 35971004 Open Access

Divakaruni AS, Jastroch M (2022) Nat Metab

Abstract: Measurement of oxygen consumption is a powerful and uniquely informative experimental technique. It can help identify mitochondrial mechanisms of action following pharmacologic and genetic interventions, and characterize energy metabolism in physiology and disease. The conceptual and practical benefits of respirometry have made it a frontline technique to understand how mitochondrial function can interface with-and in some cases control-cell physiology. Nonetheless, an appreciation of the complexity and challenges involved with such measurements is required to avoid common experimental and analytical pitfalls. Here we provide a practical guide to oxygen consumption measurements covering the selection of experimental models and instrumentation, as well as recommendations for the collection, interpretation and normalization of data. These guidelines are provided with the intention of aiding experimental design and enhancing the overall reputability, transparency and reliability of oxygen consumption measurements.

β€’ Bioblast editor: Gnaiger E

Correction

Under 'Chamber-based platinum electrode', Oroboros Instruments is listed as a 'vendor'. The Oroboros O2k uses not platinum but gold as the cathode of the polarographic oxygen sensor.