Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Talk:MiPNet17.05 O2k-Fluo LED2-Module"

From Bioblast
(Blanked the page)
Line 1: Line 1:
=Titration and settings=
* 10 µl  AmR - 5 µM final conc. in chamber
* 4µl HRP (500U/ml) - 1 U/ml final conc. in chamber* light intensity of LEDs: 1 mA ( DatLab: Amp polaristion voltage 100 mV) / Gain 1000
* H<sub>2</sub>O<sub>2</sub> calibration with a''' 15.8 µM H<sub>2</sub>O<sub>2</sub>''' solution - '''2012-06-20'''
:: '''commercial stock:''' 880 mM
:: dilution 1 (1:166.67): 150 µl in 25 ml H<sub>2</sub>O = 5.28 mM
:: dilution 2 (1:333.33): 150 µl in 50 ml H<sub>2</sub>O + 500 µl 1 mM HCl = '''15.8 µM'''
:: → '''calibration:''' 2 * 5 µl in 2 ml chamber = 2* 0.0395 µM = total 0.079 µM (= 79 nM)


* new H<sub>2</sub>O<sub>2</sub> calibration with a '''28.16 µM''' H<sub>2</sub>O<sub>2</sub> solution, valid from from '''2012-06-22''':
:: '''commercial stock:''' 880 mM
:: dilution 1 (1:125): 200 µl in 25 ml H<sub>2</sub>O = 7.04 mM
:: dilution 2 (1:250): 200 µl in 50 ml H<sub>2</sub>O + 500 µl 1 mM HCl = '''28.16 µM'''
:: → '''calibration''': 2 * 5 µl in 2 ml chamber = 2* 0.0704 µM = total 0.1408 µM (= 140.8 nM)

Revision as of 23:10, 23 July 2015