Difference between revisions of "TIP2k syringe blocked"
From Bioblast
(Created page with "==Problem== *TIP syringe is harder to operate than usually. ==Solution== Start immediately with extensive cleaning procedures [http://www.oroboros.at/index.php?id=o2k-tip2k-ma...") Β |
Β |
||
(8 intermediate revisions by 5 users not shown) | |||
Line 1: | Line 1: | ||
== | {{Technical support}} | ||
__TOC__ | |||
{{MitoPedia without banner | |||
|abbr= | |||
|description=When the '''TIP2k syringe is blocked''', it must not be used with the TIP2k, and specific cleaning instructions should be followed. | |||
|info=[https://wiki.oroboros.at/images/f/f5/MiPNet12.10_TIP2k-Manual.pdf MiPNet12.10 TIP2k-manual.pdf] | |||
}} | |||
:::# Start immediately with extensive cleaning procedures. | |||
:::# Suitable solvents for extensive cleaning procedures depend on the substances used. Chloroform was successful for some fatty acids. | |||
:::# An ultrasonic bath may be helpful in cleaning partially blocked syringes. | |||
::::* Once a needle is totally blocked it is usually impossible (with exceptions) to clean it. | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Β | |||
Latest revision as of 13:05, 8 November 2019
TIP2k syringe blocked |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
When the TIP2k syringe is blocked, it must not be used with the TIP2k, and specific cleaning instructions should be followed.
Reference: MiPNet12.10 TIP2k-manual.pdf
- Start immediately with extensive cleaning procedures.
- Suitable solvents for extensive cleaning procedures depend on the substances used. Chloroform was successful for some fatty acids.
- An ultrasonic bath may be helpful in cleaning partially blocked syringes.
- Once a needle is totally blocked it is usually impossible (with exceptions) to clean it.