Difference between revisions of "Search for defective O2k components"
From Bioblast
(Created page with "{{Technical support}} {{MitoPedia without banner |abbr= |description=The 2-chamber design of the O2k helps to '''search for defective O2k components''', by switching component...") ย |
|||
Line 5: | Line 5: | ||
}} | }} | ||
:::: The '''two chamber design''' of the O'''2k''' has many benefits, including advantages in '''trouble shooting'''. By switching components between sides A and B of the O2k, it is usually possible to locate an instrumental problem, finding selectively a specific defective component of the O2k. This component can then be serviced by the user or replaced by Oroboros without the need to ship the entire O2k.ย ย ย ย | :::: The '''two chamber design''' of the O'''2k''' has many benefits, including advantages in '''trouble shooting'''. By switching components between sides A and B of the O2k, it is usually possible to locate an instrumental problem, finding selectively a specific defective component of the O2k. This component can then be serviced by the user or replaced by Oroboros without the need to ship the entire O2k.ย ย ย ย | ||
ย | [For more information including how to disassemble the OroboPos https://wiki.oroboros.at/index.php/Oxygen_sensor_test#Troubleshooting] | ||
ย | |||
{{MitoPedia O2k and high-resolution respirometry | {{MitoPedia O2k and high-resolution respirometry | ||
|mitopedia O2k and high-resolution respirometry=O2k hardware | |mitopedia O2k and high-resolution respirometry=O2k hardware | ||
}} | }} |
Revision as of 12:09, 12 December 2019
Search for defective O2k components |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
The 2-chamber design of the O2k helps to search for defective O2k components, by switching components linked to O2k chambers A and B between sides A and B.
- The two chamber design of the O2k has many benefits, including advantages in trouble shooting. By switching components between sides A and B of the O2k, it is usually possible to locate an instrumental problem, finding selectively a specific defective component of the O2k. This component can then be serviced by the user or replaced by Oroboros without the need to ship the entire O2k.
[For more information including how to disassemble the OroboPos https://wiki.oroboros.at/index.php/Oxygen_sensor_test#Troubleshooting]
MitoPedia O2k and high-resolution respirometry: O2k hardware