Polarization voltage

From Bioblast


high-resolution terminology - matching measurements at high-resolution


Polarization voltage

Description

A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.

Abbreviation: U



MitoPedia methods: Respirometry, "DatLab" is not in the list (Respirometry, Fluorometry, Spectrophotometry) of allowed values for the "MitoPedia method" property. DatLab 

Template:MitoPedia SUIT


DatLab

The polarization voltage is set in DatLab in the "Oxygraph-2k" \ "O2k-Control" window (MiPNet19.18A O2k-Start). The default value is 800 mV.

... more about "Polarization voltage"
A polarization voltage of 600 mV to … A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.be confused with the polarization voltage. +
Respirometry +
O2 sensor +  and DatLab +