Description
A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.
Abbreviation: U
MitoPedia methods:
Respirometry
High-resolution respirometry
The polarization voltage is set in DatLab in the "Oxygraph-2k" \ "Control" window (MiPNet12.06). The default value is 800 mV.
... more about "Polarization voltage"
A polarization voltage of 600 mV to β¦ A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.be confused with the polarization voltage. +