Description
A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.
Abbreviation: U
MitoPedia O2k and high-resolution respirometry:
O2k hardware,
DatLab
DatLab
The polarization voltage is set in DatLab in the "Oxygraph-2k" \ "O2k control" window (MiPNet19.18A O2k-Start). The default value is 800 mV.