POS (technical service)

From Bioblast

As the heart of the OROBOROS Oxygraph the Polarographic oxygen sensor is of course the prime suspect when any kind of troubles are encountered. However, frequently problems can be traced either to inadequate software settings or to other hardware components (like the sensor connector). Therefore it is important to confirm the problem doing a Sensor test and to localize the problem.

Possible Problems

Problems that require a full Sensor test and Localization of the Problem

  • unstable O2 signal
  • unstable O2 flux
  • high O2 signal at air saturation
  • high O2 signal at zero oxygen

Special Cases: no further localization necessary

Performing the Sensor Test

Confirm the problem by doing a Sensor test in the absence of biological material

Option 1.A: No problem was visible in the Sensor Test

check the following DatLab settings

If the problem is not visible when observing the Raw signal but is visible when observing the calibrated oxygen signal, then there is probable a problem with the POS calibration.

Option 1.B: The problem was visible in the Sensor Test

  • proceed to Localizing the Problem

Localizing the Problem

Locate the Problem

Option 2.A: The problem could be localized on the POS

Solutions 2.A Follow the instructions for oxygen sensor service [MiPNet08.04]. Use this opportunity to apply contact oil to the gold pin and thread connecting POS and sensor connector. After a sensor service or after applying anew membrane allow the oxygraph to run over night to achieve a stable signal before repeating the sensor test. If all service precautions fail to solve the problem, the POS has to be sent to us for further service (a recovery can not be guaranteed), or a new sensor is to be applied.

Option 2.B: The problem was not localized on the POS

Solutions 2.B If the problem was localized on an other component follow the instructions for this component. If you could not locate the problem contact OROBOROS Instruments.





Template:Technical service