Line 15: Line 15:
:::* Version 08 (2020-10-17) Section '''5.4. O2k-chamber volume calibration'''
:::* Version 08 (2020-10-17) Section '''5.4. O2k-chamber volume calibration'''
::::# Instructions are added for the 0.5-mL O2k-chamber volume calibration ([[O2k-sV-Module]]).
::::# Instructions are added for the 0.5-mL O2k-chamber volume calibration ([[O2k-sV-Module]]).
::::# Volume calibration of 2-mL O2k-chamber: Fill open chamber with 2.07 mL water (not 2.085 or 2.10 mL). The volume of the stopper capillary is 0.07 mL (not 0.085 mL).
::::# Volume calibration of 2-mL O2k-chamber: Fill open chamber with 2.07 mL water (not 2.085 or 2.10 mL). The volume of the stopper capillary is 0.065 mL (not 0.085 mL).





Revision as of 10:36, 18 November 2020


                  


O2k-Open Support

MiPNet22.11 O2k-FluoRespirometer manual


Publications in the MiPMap
O2k-Manual
O2k-FluoRespirometer manual.

» Bioblast pdf» Versions

Oroboros (2020-10-17) Mitochondr Physiol Network

Abstract: Gnaiger E, Mészáros A, Gradl L, Gradl P (2020) O2k-FluoRespirometer manual. Mitochondr Physiol Network 22.11(08):1-17.

O2k-technical support and open innovation
Open the pdf document above.
» Current O2k-series: NextGen-O2k Series XB and O2k Series J
» Current software versions DatLab 8.0: MitoPedia: DatLab
» Product: DatLab 7, O2k-FluoRespirometer, O2k-Catalogue


O2k-Network Lab: AT_Innsbruck_Oroboros

Important updates

  • Version 08 (2020-10-17) Section 5.4. O2k-chamber volume calibration
  1. Instructions are added for the 0.5-mL O2k-chamber volume calibration (O2k-sV-Module).
  2. Volume calibration of 2-mL O2k-chamber: Fill open chamber with 2.07 mL water (not 2.085 or 2.10 mL). The volume of the stopper capillary is 0.065 mL (not 0.085 mL).


Keywords


Questions.jpg


Click to expand or collaps



Labels: MiParea: Instruments;methods 





HRR: O2k-Manual, Oxygraph-2k 