Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

MiPNet09.12 O2k-Titrations

From Bioblast
Revision as of 10:33, 13 September 2019 by Doerrier Carolina (talk | contribs)
Publications in the MiPMap
O2k-Protocols
O2k manual titrations: SUIT protocols with mitochondrial preparations.

» Bioblast pdf »Versions

Oroboros (2019-09-13) Mitochondr Physiol Network

Abstract: Oroboros (2019) O2k manual titrations: SUIT protocols with mitochondrial preparations. Mitochondr Physiol Network 09.12(18): 1-2.

A list of substrates, uncouplers, and inhibitors used in SUIT protocols in high-resolution respirometry, with information on abbreviations, stock concentrations, final experimental concentrations, volumes titrated into the 2 mL O2k-chamber, and syringes used for manual titrations.

» O2k-Protocols
» Product: Oroboros O2k, O2k-Catalogue


O2k-Network Lab: AT_Innsbruck_Oroboros


Labels: MiParea: Instruments;methods 




Regulation: ADP, Inhibitor, Substrate, Uncoupler, Amino acid 


HRR: O2k-Protocol 

O2k-chemicals and media 


Note: Titration of chemicals in respirometry

Overshoots or short interruptions of the flux after the injection of chemicals do not have a meaning in respirometric measurements. A disturbance of the signal is only observed due to the injected chemical and does not represent mitochondrial respiration. With the addition of a chemical solved in ethanol a little bit of oxygen is always introduced, since ethanol has higher oxygen solubility than water.


Further information

» MitoPedia: SUIT - Substrate-uncoupler-inhibitor-titration (SUIT) protocols for respirometry
» MiPNet03.02 - Chemicals and media for respirometry
» MiPNet19.14 - SOP for manual O2k-titrations with Hamilton microsyringes.