Difference between revisions of "Filter-Cap"
From Bioblast
Tindle Lisa (talk | contribs) |
Tindle Lisa (talk | contribs) m ย |
||
Line 3: | Line 3: | ||
{{MitoPedia without banner | {{MitoPedia without banner | ||
|abbr= | |abbr= | ||
|description=[[Image:Filter-Cap.JPG|180px|right]]'''Filter-Cap''': O2k-Fluo LED2-Module (O2k-Series D to G) sensors ([[Fluorescence-Sensor Green]] and [[Fluorescence-Sensor Blue]]) and O2k-FluoRespirometer (O2k-Series H) sensors ([[Smart Fluo-Sensor Green]] and [[Smart Fluo-Sensor Blue]]) are equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode. | |description=[[Image:Filter-Cap.JPG|180px|right]]'''Filter-Cap''': O2k-Fluo LED2-Module (O2k-Series D to G) sensors ([[Fluorescence-Sensor Green]] and [[Fluorescence-Sensor Blue]]) and O2k-FluoRespirometer (O2k-Series H to I) sensors ([[Smart Fluo-Sensor Green]] and [[Smart Fluo-Sensor Blue]]) are equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode. | ||
|info= [https://www.oroboros.at/index.php/product/filter-cap/ '''Product details and purchase information''']}} | |info= [https://www.oroboros.at/index.php/product/filter-cap/ '''Product details and purchase information''']}} | ||
Line 14: | Line 14: | ||
=== Mounting a Filter-Cap === | === Mounting a Filter-Cap === | ||
:::* '''Applies to''': O2k-Fluo LED2-Module (O2k-Series D to G) Fluorescence Sensor; O2k-FluoRespirometer (O2k-Series H) with Smart Fluo-Sensors | :::* '''Applies to''': O2k-Fluo LED2-Module (O2k-Series D to G) Fluorescence Sensor; O2k-FluoRespirometer (O2k-Series H to I) with Smart Fluo-Sensors | ||
:::* Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. ย | :::* Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements. ย | ||
:::* Remove all filters and store them in the filter box labeled for this Filter Set. ย | :::* Remove all filters and store them in the filter box labeled for this Filter Set. ย |
Latest revision as of 09:40, 6 May 2021
Filter-Cap |
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Description
Filter-Cap: O2k-Fluo LED2-Module (O2k-Series D to G) sensors (Fluorescence-Sensor Green and Fluorescence-Sensor Blue) and O2k-FluoRespirometer (O2k-Series H to I) sensors (Smart Fluo-Sensor Green and Smart Fluo-Sensor Blue) are equipped with a removable Filter-Cap for exchange of optical filters for the optical pathways from the LED to the sample and from the sample to the photodiode.
Reference: Product details and purchase information
Technical Support
Selecting a Filter Set
- The Filter-Cap of each sensor can be removed for application of various filter combinations on the same optical sensor.
- Selection of a Filter Set: >> Application specific settings.
Mounting a Filter-Cap
- Applies to: O2k-Fluo LED2-Module (O2k-Series D to G) Fluorescence Sensor; O2k-FluoRespirometer (O2k-Series H to I) with Smart Fluo-Sensors
- Dismounting: Pull the Filter-Cap straight from the sensor. The Filter-Cap Guide prevents rotational movements.
- Remove all filters and store them in the filter box labeled for this Filter Set.
- Insert the filters form the selected Filter Set: The round filters of each Filter Set fit to the round window of the Filter-Cap and cover the LED, the rectangular filters fit into the rectangular window of the Filter-Cap and cover the photodiode.
- Mounting: Hold sensor and filter cap in a vertical position above you. Align the Filter-Cap with the Filter-Cap Guide (small steel rod) protruding from the sensor. Press the Filter-Cap onto the sensor without rotational movements.
- O2k-Guide: Setup - next step Connect Fluorescence-Sensor to O2k
Additional resources
MitoPedia O2k and high-resolution respirometry:
O2k hardware,
O2k-Open Support,
O2k-Respirometry,
O2k-FluoRespirometry