Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Cyanide"

From Bioblast
(formatting)
(The product number belongs to Sigma, not Fluka. Corrected that)
Line 21: Line 21:
=== Preparation of KCN stock ===
=== Preparation of KCN stock ===


:::: '''Kcn'''(KCN): Fluka 60178, 100 g, store at R.T.; M = 65.12 g·mol<sup>-1</sup>.
:::: '''Kcn'''(KCN): Sigma 60178, 100 g, store at R.T.; M = 65.12 g·mol<sup>-1</sup>.
:::: '''Caution:''' In April 2019 a change in the procedure for preparing the KCN was done that recommended to prepare 20 mM KCN stock and titrate 100 µL. 2021-01-22: after discussions on KCN toxicity we recommend to prepare of 0.5 M KCN stock solutions.
:::: '''Caution:''' In April 2019 a change in the procedure for preparing the KCN was done that recommended to prepare 20 mM KCN stock and titrate 100 µL. 2021-01-22: after discussions on KCN toxicity we recommend to prepare of 0.5 M KCN stock solutions.



Revision as of 12:14, 16 March 2021


high-resolution terminology - matching measurements at high-resolution


Cyanide

Description

Cyanide (usually added as KCN) is a competitive inhibitor of cytochcrome c oxidase (CIV). Inhibition is reversed by pyruvate and high oxygen levels.

Abbreviation: KCN

Reference: MiPNet09.12



MitoPedia topics: Inhibitor 

Application in HRR

Last update 2021-01-22

Inhibitor titrations, CIV-assay, alternative oxidase and ROX


Preparation of KCN stock

Kcn(KCN): Sigma 60178, 100 g, store at R.T.; M = 65.12 g·mol-1.
Caution: In April 2019 a change in the procedure for preparing the KCN was done that recommended to prepare 20 mM KCN stock and titrate 100 µL. 2021-01-22: after discussions on KCN toxicity we recommend to prepare of 0.5 M KCN stock solutions.
Caution: KCN is photosensitive and hygroscopic; HCN is volatile and very toxic!


Preparation of 0.5 M KCN stock solution (dissolved in H2O):
  1. Weigh 32.6 mg of KCN.
  2. Dissolve in 1 ml H2O.
  3. The pH of the solution is basic. Since the titration amount is low and HCN may evaporate during pH adjustment with HCl, the pH is not adjusted.
  4. Divide into 0.2 mL portions
  5. Store at -20 °C.
» O2k manual titrations MiPNet09.12 O2k-Titrations
  • Titration volume (2-mL O2k-chamber): 4 µL using a 10 µL Hamilton syringe.
  • Titration volume (0.5-mL O2k-chamber): 1 µL using a 10 µL Hamilton syringe.
  • Final concentration: 1 mM.



MitoPedia topics: Inhibitor