MiPNet09.12 Manual titrations of SUIT chemicals

From Bioblast
(Redirected from MiPNet09.12)

                



MiPNet09.12 Manual titrations of SUIT chemicals

Publications in the MiPMap
O2k-Protocols
Manual titrations of SUIT chemicals with 2.0 mL and 0.5 mL chambers of the Oroboros Bioenergetics Analyzer.

Β» Bioblast pdf Β»Versions

Oroboros (2024-11-25) Mitochondr Physiol Network

Abstract: Oroboros (2024) Manual titrations of SUIT chemicals with 2.0 mL and 0.5 mL chambers of the Oroboros Bioenergetics Analyzer. Mitochondr Physiol Network 09.12(21):1-4.

Oroboros technical support and open innovation
Open the pdf document above.
Β» Current Oroboros series: NextGen-O2k Series XB and O2k Series J
Β» Current software versions DatLab 8.2: MitoPedia: DatLab
Β» O2k-Protocols
Β» Product: Oroboros


β€’ O2k-Network Lab: AT_Innsbruck_Oroboros


Note

Oligomycin: titrations should be performed by 1-2 Β΅L steps (i.e. 5-10 nM experimental concentration of each step). Careful titration of oligomycin is necessary to determine the optimal inhibitory concentration (from 10 to 40 nM) as elaborated in Vernerova et al 2021 and Zdrazilova et al 2022 and to prevent inhibition of ET respiration as detailed in Ruas et al (2016) and Doerrier et al 2018.


Labels: MiParea: Instruments;methods 




Regulation: ADP, Inhibitor, Substrate, Uncoupler, Amino acid 


HRR: O2k-Protocol 

O2k-chemicals and media, ADP 


Cookies help us deliver our services. By using our services, you agree to our use of cookies.