The gain is an amplification factor applied to an input signal to increase the output signal.
MitoPedia Oroboros and high-resolution respirometry:
Oroboros Open Support
Gain: O2 channel of the O2k
- For most applications between 25 and 37 Β°C the pre-set gain of 1 (default) is adequate and does not have to be changed.
- The gain of the signal of the O2 channel can be adjusted in DatLab by selecting [Oroboros O2k] / [O2k Control] / Tab [Oxygen, O2] from the menu. When the gain setting is too high for any given signal, it will reach a constant maximum value of 10 V (often displayed as 9.99 V) and the displayed slope will therefore apparently be constant at zero.
- For each gain setting a separate calibration of the oxygen signal is required. Therefore, after changing the gain you should not only perform a calibration at air saturation (as always before an experiment) but also a new zero calibration, see Air calibration. If for any reason you frequently switch between gains, the easiest way is to store the calibrations in different DatLab files and use the "read from file" function in the calibration window. Alternatively, values can be entered manually.
- Oxygen signals that have been obtained after proper calibration are independent of the gain setting. Therefore oxygen concentrations (and fluxes) that have been recorded at different gain settings can of course be used together, as long as the proper calibration is applied for these files.
- See also Β»Oxygen signal