Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Talk:MiPNet17.08 IOC68 Schroecken AT

From Bioblast

Please add you Feedback here.


IOC68 Workshop Experiments

2012-04-14 Demo Experiment

  • Oxygraph AB (in both chambers the same SUIT protocol was applied)
  • DatLab File 2012-04-14_AB-02
  • mouse heart homogenate prepared with PBI Shredder (position 1 for 10 seconds, position 2 for 5 seconds) and the homogenate was completely washed out of the Shredder tube with 5 ml MiR05
  • 2.3 mg tissue per chamber = 1.15 mg / ml
  • Flux derivation = 20
  • oxygen regime 320 to 160 (reoxygenation with pure oxygen at open chamber position)
  • respiration medium: MiR05 (solubility factor = 0.92)
  • SUIT-Protocol: PMG + AmR + HRP + H2O2 + H2O2 + D2 + D3 + S + FCCP + Rot + H2O2 + H2O2 + Mna + Ama
  • the homogenate was added, the oxygen concentration was increased till 320 nmol/ml and then the chamber was closed

Notes

  • there is a dependence of oxygen with homogenate but not that strong as with fibres (decrease of oxygen flux with decrease of oxygen concentration)
  • stimulation of respiration with ADP and succinate
  • reoxygenation with pure oxygen causes an increase in peroxide production
  • no big effect of FCCP on oxygen flux but a decrease in peroxide production occurred
  • oxygen dependence visible in the decline of oxygen flux and NO signal
  • rotenone increased the peroxide production
  • analysis of the experiments is difficult because of O2 dependence