Polarization voltage: Difference between revisions
From Bioblast
No edit summary |
No edit summary |
||
Line 5: | Line 5: | ||
{{MitoPedia concepts}} | {{MitoPedia concepts}} | ||
{{MitoPedia methods | {{MitoPedia methods | ||
|mitopedia method=Respirometry | |mitopedia method=Respirometry | ||
}} | }} | ||
{{MitoPedia O2k and high-resolution respirometry | |||
|mitopedia O2k and high-resolution respirometry=DatLab | |||
}} | |||
{{MitoPedia topics}} | |||
{{MitoPedia SUIT}} | {{MitoPedia SUIT}} | ||
== DatLab == | == DatLab == | ||
The polarization voltage is set in [[DatLab]] in the "Oxygraph-2k" \ "O2k-Control" window ([[MiPNet19.18A O2k-Start]]). The default value is 800 mV. | The polarization voltage is set in [[DatLab]] in the "Oxygraph-2k" \ "O2k-Control" window ([[MiPNet19.18A O2k-Start]]). The default value is 800 mV. |
Revision as of 10:55, 8 February 2016
Description
A polarization voltage of 600 mV to 800 mV is applied between anode and cathode of the polarographic oxygen sensor, resulting in a current when oxygen is consumed. The current is converted by the electronics to a voltage (raw signal) which must not be confused with the polarization voltage.
Abbreviation: U
MitoPedia methods: Respirometry
MitoPedia O2k and high-resolution respirometry:
DatLab
DatLab
The polarization voltage is set in DatLab in the "Oxygraph-2k" \ "O2k-Control" window (MiPNet19.18A O2k-Start). The default value is 800 mV.