Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "MiPNet20.14 AmplexRed H2O2-production"

From Bioblast
 
(32 intermediate revisions by 8 users not shown)
Line 1: Line 1:
{{OROBOROS header page name}}
{{Publication
{{Publication
|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] O2k-Fluorometry: HRR and simultaneous determination of H<sub>2</sub>O<sub>2</sub> production with Amplex Red. [[Media: MiPNet20.14_AmplexRed H2O2-production.pdf |»Bioblast pdf«]]
|title=[[Image:O2k-Protocols.jpg|right|80px|link=O2k-Protocols|O2k-Protocols]] O2k-FluoRespirometry: HRR and simultaneous determination of H<sub>2</sub>O<sub>2</sub> production with Amplex UltraRed.
|info=[http://wiki.oroboros.at/index.php/File:MiPNet20.14_AmplexRed H2O2-production.pdf Versions]
|info=[[File:PDF.jpg|100px|link=http://wiki.oroboros.at/images/5/57/MiPNet20.14_AmplexRed_H2O2-production.pdf|Bioblast pdf]] »[http://wiki.oroboros.at/index.php/File:MiPNet20.14_AmplexRed_H2O2-production.pdf Versions]
|authors=OROBOROS
|authors=Oroboros
|year=2015-06-08
|year=2019-06-24
|journal=Mitochondr Physiol Network
|journal=Mitochondr Physiol Network
|abstract=Krumschnabel G, Fontana-Ayoub M, Fasching M, Gnaiger E (2015) O2k-Fluorometry: HRR and simultaneous determination of H<sub>2</sub>O<sub>2</sub> production with Amplex Red. Mitochondr Physiol Network 20.14(01):1-6.
|abstract=Krumschnabel G, Fontana-Ayoub M, Fasching M, Gnaiger E (2019) O2k-FluoRespirometry: HRR and simultaneous determination of H<sub>2</sub>O<sub>2</sub> production with Amplex UltraRed. Mitochondr Physiol Network 20.14(04):1-6.  
 
{{MiPNet pdf page linking to MitoPedia}}
[[High-resolution respirometry]] [1] is extended by O2k-MultiSensor modules allowing the simultaneous measurement of oxygen consumption and one additional parameter (mitochondrial membrane potential, pH, Ca<sup>2+</sup>, or NO concentration [2-5]). Fluorometric methods are available for a wide range of analytical parameters of major interest in mitochochondrial physiology: H<sub>2</sub>O<sub>2</sub> production, mitochondrial membrane potential, intracellular pH, Ca<sup>2+</sup>, Mg<sup>2+</sup>, and NADH levels.
|mipnetlab=AT_Innsbruck_Oroboros
Advantages of the simultaneous measurement of additional parameters in a single respirometric chamber:
# Quality control of preparations by respirometric performance.
# Objective exclusion criteria for protocol related problems, e.g. overtitration of uncouplers, side effects of TPP<sup>+</sup> or fluorophores.
# Elimination of artifacts of normalization for a mitochondrial marker when combining two methods in a single chamber.
# Additional information is acquired for a limited amount of sample.
# Oxygen dependence of parameters can be studied.
In the publication (Krumschnabel et al 2015) we introduce the O2k-Fluo LED2-Module, describe one selected stage in its development, and show its application for the simultaneous measurement of respiration and H<sub>2</sub>O<sub>2</sub> production of mouse brain mitochondria.
 
'''Reference''': Krumschnabel G, Fontana-Ayoub M, Sumbalova Z, Heidler J, Gauper K, Fasching M, Gnaiger E (2015) Simultaneous high-resolution measurement of mitochondrial respiration and hydrogen peroxide production. Methods Mol Biol 1264:245-61. [[Krumschnabel 2015 Methods Mol Biol |»Bioblast pdf«]]
|mipnetlab=AT_Innsbruck_OROBOROS
}}
}}
{{Labeling
{{Labeling
Line 25: Line 15:
|tissues=Heart
|tissues=Heart
|preparations=Isolated mitochondria
|preparations=Isolated mitochondria
|couplingstates=LEAK, OXPHOS, ETS
|couplingstates=LEAK, OXPHOS, ET
|substratestates=CII
|pathways=S
|instruments=Oxygraph-2k, O2k-Fluorometer, Protocol
|instruments=Oxygraph-2k, O2k-Fluorometer, O2k-Protocol
|additional=O2k-Demo, O2k-MultiSensor
|additional=O2k-Demo, O2k-MultiSensor, MitoFitPublication, SUIT-013, SUIT-013 AmR ce D023, SUIT-009, SUIT-009 AmR mt D021, SUIT-009 AmR pce D019, SUIT-018, SUIT-018 AmR mt D031, SUIT-006 AmR mt D048, SUIT-003 AmR ce D017, SUIT-003 AmR ce D058, H2O2
}}
}}
== Applications with O2k-Fluorometry ==
The Amplex Red method<ref>[[Mohanty 1997 J Immunol Methods]]</ref> is used with the [[O2k-Fluo LED2-Module]] selecting the [[Fluorescence-Sensor Green]]. Use [[Stopper\black PEEK\conical Shaft\central Port |black stoppers]] with [[Cover-Slip\black |black cover-slips]] to exclude disturbances by external light sources.
The fluorescence sensors are inserted though the windows of the O2k-Chambers<ref>[[Hickey 2012 J Comp Physiol B]]</ref>. The O2k has also been coupled to a fluorescence spectrophotometer with a light guide inserted through the [[Stopper\black PEEK\conical Shaft\central+side 2.5 mm Port|black PEEK stopper]].<ref> [[Anderson 2011 Am J Physiol Heart Circ Physiol]])</ref>
== Preparation of Amplex® UltraRed solutions ==
Source: Life Technologies (former Invitrogen) A36006
See the manual provided by the producer: [http://tools.lifetechnologies.com/content/sfs/manuals/mp36006.pdf Manual Amplex UltraRed].
=== Storage solution = stock solution ===
Prepare a 10 mM storage solution of Amplex® UltraRed reagent (AmR), which will also serve as stock solution, by adding 340 μL of fresh, high‑quality DMSO (Sigma D8418) to one commercial vial of Amplex® UltraRed reagent (1 mg). Vortex well to dissolve. Protect AmR from light and moisture. The storage solution should then be divided into 10 µl aliquots and stored in the dark, protected from moisture at –20°C for future use. When stored properly, the 10 mM storage solution is stable for at least 6 months.
Note: The preparation of the 10 mM storage solution follows the procedure suggested by the manufacturer. The manufacturer states only an approximate molecular weight for the Amplex® UltraRed formulation and does not publish details how the Amplex® UltraRed formulation deviates from the substance 10-acetyl-3,7-dihydroxyphenoxazine (CAS# 119171-73-2), known as Amplex® Red. 
=== Final AmR concentration ===
* Titration into into 2 ml O2k-chamber: '''2 µl of 10 mM AmR stock solution''', final concentration of '''10 µM AmR'''.
You may add 1 µl of the 10 mM stock solution for brief measurements to obtain a final concentration of 5 µM. Optimize the final AmR concentration according to respiration media, sample type and concentration, and experimental protocol. These determine the total AmR consumption by H<sub>2</sub>O<sub>2</sub> production during the experiment (check by initial and final H<sub>2</sub>O<sub>2</sub> titrations for calibration). The ''final'' concentration of AmR becomes diminished during an experiment due to AmR consumption and dilution by titrations in a [[SUIT protocol]]. Check for any potential effects of the exprimental AmR concentration on respiratory activity of cells or mt-preparations, by titration of AmR after a specific respiratory state has been reached in a control experiment.
==== Horse radish peroxidase ====
* Titration into into 2 ml O2k-chamber: '''4 µl of HRP stock solution''', final concentration '''1 U/ml'''.
[[Horseradish_peroxidase|Horse radish peroxidase]] (Sigma-Aldrich P 8250 - 5 kU): Prepare a stock solution with 500 U HRP/ml in MiR05 or MiR05Cr; the solution can be used as storage solution at -20 °C.
       
==== Superoxide dismutase ====
[[Superoxide dismutase]] (SOD) is optionally included to generate H<sub>2</sub>O<sub>2</sub> from superoxide. Results may be compared with and without SOD for evaluation of the contribution of superoxide not endogenously dismutated with the formation of peroxide.
(Sigma-Aldrich S 8409-15KU): Depending on the batch the SOD preparation may contain a specific activity of 2,000–6,000 units/mg protein. We recommend to use the enzyme at 5 U/ml, the final volume to be added to the respiratory chamber has thus to be adjusted accordingly.
   
== Calibration with H<sub>2</sub>O<sub>2</sub> ==
Calibration standards of H<sub>2</sub>O<sub>2</sub>: Commercial solution (Sigma-Aldrich 323381 - 25 ml  hydrogen peroxide solution 3 wt. %; stabilized with acetanilide, c. 200 ppm) = 880 mM H<sub>2</sub>O<sub>2</sub>.
# To prepare a HCl stock solution of 10 µM HCl, take 1 ml of 1 mM HCl storage solution an fill up to 100 ml with distilled water.
# H<sub>2</sub>O<sub>2</sub> dilution 1 (1:88): add 284 µl of a commercial solution of 3 wt.% H<sub>2</sub>O<sub>2</sub> to 20 ml of 10 µM HCl and fill up to a final volume of 25 ml with 10 µM HCl, to obtain a 10 mM H<sub>2</sub>O<sub>2</sub> solution.
# H<sub>2</sub>O<sub>2</sub> dilution 2 (1:250): dilute 200 µl of 10 mM H<sub>2</sub>O<sub>2</sub> solution with 10 µM HCl to a volume of 50 ml to obtain the H<sub>2</sub>O<sub>2</sub> stock solution of 40 µM.
* Titration into into 2 ml O2k-Chamber: '''5 µl of the 40 µM H<sub>2</sub>O<sub>2</sub> stock''', step increase of '''0.1 µM  H<sub>2</sub>O<sub>2</sub>'''. One to three subsequent steps.
Titrations are peformed after addition of respiration medium, HRP and Amplex UltraRed to the O2k chamber. We suggest to perform calibrations at the beginning (sample already present), intermittently at various respiratory states, and near the end of an experiment (but before cytochrome ''c'', catalase or other substances are added that are interfer with the H<sub>2</sub>O<sub>2</sub> measurement).
DatLab supports automatic H<sub>2</sub>O<sub>2</sub> calibration by calculating the calibration parameters by linear regression and graphical display of the calibration regression:
>> [[MiPNet19.18D_O2k-Calibration]].
Since part of the signal change might be caused by an ongoing  H<sub>2</sub>O<sub>2</sub> production of the biological sample the calibration can be corrected by including the slope in the calculation.
'''DatLab 6'''
The slope used for the correction is calculated from a linear regression of the signal vs time for the selected marks. This eliminates the distortion of the slope by injection artifacts after the Amp signal is already stable.
* Select/deselect slopes to be include in the calculation by ticking the check boxes on the right of the slope values. If you select more than one slope the weighted mean of all values is used for calculation.
== Substances incompatible with the Amplex Red method ==
The following substances/ classes of substances are strictly incompatible with the Amplex Red method for theoretical reasons:
* Strongly redox active substances, e.g. cytochrome ''c'', TMPD/Ascorbate
* Inhibitors of [[horseradish peroxidase]], e.g. cyanide, azide; more information: [[horseradish peroxidase]]
* Catalase and other substances consuming or scavanging H<sub>2</sub>O<sub>2</sub>. The effect of substances in the medium that consume H<sub>2</sub>O<sub>2</sub> slowly is taken account of by the calibration procedure. However, such substances decrease the sensitivity of the method. Note that catalase can be a valuable tool for [[Amplex_red#Checking_for_artifacts|checking for artifacts]]. 
The effect of other substances should be checked by blank experiments, including comparing the sensitivity (result of calibration) before and  after injecting the substances.
In  preliminary experiments OROBOROS INSTRUMENTS evaluated small amounts (as typically used in SUIT protocols) of the following substances as '''compatible with the method''':
DMSO,  ethanol, [[malate]], [[glutamate]], [[pyruvate]] (a strong scavanger of H<sub>2</sub>O<sub>2</sub>), [[succinate]],  ([[ADP]] + Mg<sup>2+</sup>), ([[ATP]] +  Mg<sup>2+</sup>), [[rotenone]], [[FCCP]], [[oligomycin]],  [[antimycin A]], [[malonate]], [[myxothiazol]]
=== Checking for artifacts ===
The  Amplex method is based on the H<sub>2</sub>O<sub>2</sub> dependent oxidation  of AmR to resorufin by HRP. Under unfavorable conditions AmR  may be oxidized even in the absence of  H<sub>2</sub>O<sub>2</sub>. At a small rate such  an oxidation occurs in the presence of HRP even without any sample  present. The magnitude of this background signal ("drift") depends, among other things, on the light intensity used  and can therefore be minimized by using the suggested or  lower light intensity. Components of the sample may however induce a far  higher, non H<sub>2</sub>O<sub>2</sub> related rate of AmR oxidation. Therefore, especially when applying the  method on new types of samples the method should be checked for  artifacts. A few approaches are listed here:
'''Sequential  addition of HRP and AmR''': This method is particular easy to implement  if AmR and HRP have to be added to the chamber already  containing the sample anyway: First, inject Amplex Red and wait a few  minutes for flux stabilization. The Amp slope has to stay near zero. Then add HRP. The Amp slope should increase and correspond to the  H<sub>2</sub>O<sub>2</sub> production. If a  significant Amp slope is detected before the addition of HRP this  increase in fluorescence is not caused by  H<sub>2</sub>O<sub>2</sub> production. The  experiment can be continued as usual after this test. If the sample is  injected routinely into the chamber already containing AmR and HRP the  method can not be applied. In this case it is suggested to change this  sequence at least for one experiment.
'''Addition  of catalse:''' After a H<sub>2</sub>O<sub>2</sub> flux (production) is established, a high dose (e.g. 10 µl of a 280,000 U/ml stock solution) of  catalase is injected. Catalase competes with HRP for the  available H<sub>2</sub>O<sub>2</sub>. Then the apparent H<sub>2</sub>O<sub>2</sub> production (the Amp slope) should be reduced to near zero as a control for distinguishing an unspecific chemical background slope from H<sub>2</sub>O<sub>2</sub> dependent Amplex Red oxidation.
Note: The experiment cannot be continued afterwards for measurement of hydrogen peroxide production, whereas respiration can be recorded onwards.
== References ==
<references/>

Latest revision as of 12:30, 25 October 2023

                



MiPNet20.14 AmplexRed H2O2-production

Publications in the MiPMap
O2k-Protocols
O2k-FluoRespirometry: HRR and simultaneous determination of H2O2 production with Amplex UltraRed.

» Bioblast pdf »Versions

Oroboros (2019-06-24) Mitochondr Physiol Network

Abstract: Krumschnabel G, Fontana-Ayoub M, Fasching M, Gnaiger E (2019) O2k-FluoRespirometry: HRR and simultaneous determination of H2O2 production with Amplex UltraRed. Mitochondr Physiol Network 20.14(04):1-6.

O2k-technical support and open innovation
Open the pdf document above.
» Current O2k-series: NextGen-O2k Series XB and O2k Series J
» Current software versions DatLab 8.0: MitoPedia: DatLab


O2k-Network Lab: AT_Innsbruck_Oroboros


Labels: MiParea: Respiration, Instruments;methods 


Organism: Mouse  Tissue;cell: Heart  Preparation: Isolated mitochondria 


Coupling state: LEAK, OXPHOS, ET  Pathway:HRR: Oxygraph-2k, O2k-Fluorometer, O2k-Protocol 

O2k-Demo, O2k-MultiSensor, MitoFitPublication, SUIT-013, SUIT-013 AmR ce D023, SUIT-009, SUIT-009 AmR mt D021, SUIT-009 AmR pce D019, SUIT-018, SUIT-018 AmR mt D031, SUIT-006 AmR mt D048, SUIT-003 AmR ce D017, SUIT-003 AmR ce D058, H2O2